Z-1394-2015Class II

INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular...

Canon Medical Systems USA, Inc. / initiated 2015-03-11 / Terminated / root cause: design

Toshiba American Medical Systems Inc began a recall of INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular intervention with fluoroscopic and fluorographic images. For... on . FDA classified it as Class II and gives the reason as "The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1394-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular intervention with fluoroscopic and fluorographic images. For the USA and Canada the system is registered as lnfinix CF-i or lnfinix VF-i, for national regulations. INFX-8000V series includes a floor mounted multi-axis positioner.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top.
Root cause tag
design
FDA root cause
Device Design
Action
Toshiba sent a Urgent Medical Device Correction letter dated March 11, 2015, to all customers who purchased the CAT-880B catherization table for the Tosiba Infinix X-Ray Interventional System. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the attached form and fax it to (877) 349-3054 or email to [email protected]. Customers with questions were instructed to call (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968. For questions regarding this recall call 714-730-5000.
Quantity in commerce
28 units
Distribution
Nationwide Distribution
Product codes and lots
Serial No. U7E1462009, ACA1472001, UEB1352004, UFD1442017, AAA1492001, ACA1492002, ULD1412014, AHA1482001, UPB1372006, ULD1392013, U6A1472001, ULD1452016, UPB1352005, ULD1452015, UDE1352017, A4B1472002, U4A14X2001, UPB1342004, 99B1342133, 99B1342134, UFD1452018, UFD13X2015, UFD13X2016, UFD1362014, UUC1452012, A6A1492002, UUC13Z2011, AAA14X2002, A4A13Z2001.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1394-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1394-2015Class IITerminatedVoluntary: Firm initiated