Z-0005-2016Class II

Celesteion PCA-9000A/2 PET/CT System

Canon Medical Systems USA, Inc. / initiated 2015-05-08 / Terminated / root cause: design

Toshiba American Medical Systems Inc began a recall of Celesteion PCA-9000A/2 PET/CT System on . FDA classified it as Class II and gives the reason as "It was found that if specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved due to a software problem". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0005-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Celesteion PCA-9000A/2 PET/CT System
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
It was found that if specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved due to a software problem.
Root cause tag
design
FDA root cause
Device Design
Action
Planned Action by Toshiba Medical America Systems to address the defect: a. A notification will be sent to the customer with affected system in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on the system to prevent this occurrence from happening. The consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. The customer will be advised to do the following until the corrective measures have been taken, i.e., revised software installed on their systems. If you have any questions please contact your local Toshiba Representative at (800) 521-1968.
Quantity in commerce
1
Distribution
US Distribution to the state of Nevada.
Product codes and lots
Model: TSX-301 CI6A with system software V7.0 SPOO04J

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0005-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0005-2016Class IITerminatedVoluntary: Firm initiated