Z-0005-2016Class II
Celesteion PCA-9000A/2 PET/CT System
Canon Medical Systems USA, Inc. / initiated 2015-05-08 / Terminated / root cause: design
Toshiba American Medical Systems Inc began a recall of Celesteion PCA-9000A/2 PET/CT System on . FDA classified it as Class II and gives the reason as "It was found that if specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved due to a software problem". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0005-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Celesteion PCA-9000A/2 PET/CT System
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- It was found that if specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved due to a software problem.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Planned Action by Toshiba Medical America Systems to address the defect: a. A notification will be sent to the customer with affected system in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on the system to prevent this occurrence from happening. The consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. The customer will be advised to do the following until the corrective measures have been taken, i.e., revised software installed on their systems. If you have any questions please contact your local Toshiba Representative at (800) 521-1968.
- Quantity in commerce
- 1
- Distribution
- US Distribution to the state of Nevada.
- Product codes and lots
- Model: TSX-301 CI6A with system software V7.0 SPOO04J
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0005-2016 | Class II | Terminated | Voluntary: Firm initiated |