Z-0015-2016Class II

Toshiba Aquilion CT System TSX-301B

Canon Medical Systems USA, Inc. / initiated 2015-05-08 / Terminated

Toshiba American Medical Systems Inc began a recall of Toshiba Aquilion CT System TSX-301B on . FDA classified it as Class II and gives the reason as "If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0015-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba Aquilion CT System TSX-301B
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved.
FDA root cause
Radiation Control for Health and Safety Act
Action
Toshiba America Medical Planned Action to bring defect into Compliance: a. A notification will be sent to all consignees with affected systems in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on each system to prevent this occurrence from happening. Each consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. Each consignee will be advised to do the following until the corrective measures have been taken, i.e., revised software installed on their systems. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Toshiba America Medical Systems, Inc. will implement this CAP by February 1, 2016. If you have any questions please contact your local Toshiba Representative at (800) 521-1968.
Quantity in commerce
459
Distribution
US Distribution including Puerto Rico
Product codes and lots
system software V6.0 or later

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0015-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0013-2016Class IITerminatedVoluntary: Firm initiated
Z-0014-2016Class IITerminatedVoluntary: Firm initiated
Z-0015-2016Class IITerminatedVoluntary: Firm initiated
Z-0016-2016Class IITerminatedVoluntary: Firm initiated
Z-0017-2016Class IITerminatedVoluntary: Firm initiated
Z-0018-2016Class IITerminatedVoluntary: Firm initiated