INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems
Toshiba American Medical Systems Inc began a recall of INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems on . FDA classified it as Class II and gives the reason as "When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0752-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value" and it stopped displaying dose data and the dose data was lost. The following message was displayed, "Dose meter abnormal, Dose info disabled".
- Root cause tag
- software
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Toshiba's planned action to bring into compliance: 1.Affected customers will be notified using the Customer Notification Letter you provided, 2. revised system software will be installed on the systems to prevent this occurrence from happening, 3. the corrections will be made free of charge, and 4. the corrections will be completed by April 4, 2016. For further questions, please call 1- 800 521-1968.
- Quantity in commerce
- 4
- Distribution
- US Nationwide Distribution to OH and NY
- Product codes and lots
- Serial Numbers ACA1492002, AAA1492001, W4B1442020, and W4B1432018.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0752-2016 | Class II | Terminated | Voluntary: Firm initiated |