Aquilion Lightning CT System iStation cover, SID: 70000246, Item Code: TSX-0354A/2D.03

Canon Medical Systems USA, Inc. / initiated 2016-04-12 / Terminated / root cause: design

Toshiba American Medical Systems Inc began a recall of Aquilion Lightning CT System iStation cover, SID: 70000246, Item Code: TSX-0354A/2D.03 Product Usage: Aquilion Lightning is a multislice helical CT system that... on . FDA classified it as Class II and gives the reason as "Toshiba American Medical Systems, Inc. is recalling the TSX-035A because of a potential problem with the Aquilion Lightning CT System iStation cover falling off of the gantry". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1560-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Aquilion Lightning CT System iStation cover, SID: 70000246, Item Code: TSX-0354A/2D.03 Product Usage: Aquilion Lightning is a multislice helical CT system that supports whole-body imaging. The system employs Toshibas innovative dose-reduction technologies and a fast reconstruction unit designed to minimize the patient exposure dose and improve the workflow. The wide range of advanced applications is designed for operators of all levels of experience, maximizing performance. This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head. The Aquilion Lightning has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Toshiba American Medical Systems, Inc. is recalling the TSX-035A because of a potential problem with the Aquilion Lightning CT System iStation cover falling off of the gantry.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Medical Device Correction letter dated 4/12/16 was sent to a customer to bring to the customer's attention with a potential problem with the Aquilion Lightning CT System iStation cover. The letter informs the customer that the iStation cover may come off due to deficient dimensional geomtry. The letter informs the customer of the problem identified and corrective action to be taken. The customer is instructed to complete and return the attached form and fax it to (877) 349-3054. If the customer has any questions regarding the recall letter, then the customer is instructed to contact Paul Biggins at (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
Quantity in commerce
1 unit
Distribution
US in the state Nevada
Product codes and lots
Serial No. 2DA15Y2036

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1560-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1560-2016Class IITerminatedVoluntary: Firm initiated