Z-1607-2016Class II

Toshiba American Medical Systems Celesteion, Item Code: PCA-9000A/2, 9000A/3B, 9000A/2B This system is...

Canon Medical Systems USA, Inc. / initiated 2016-04-24 / Terminated / root cause: design

Toshiba American Medical Systems Inc began a recall of Toshiba American Medical Systems Celesteion, Item Code: PCA-9000A/2, 9000A/3B, 9000A/2B This system is designed to acquire CT images, PET images, and fusion... on . FDA classified it as Class II and gives the reason as "Potential degradation of detector performance". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1607-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba American Medical Systems Celesteion, Item Code: PCA-9000A/2, 9000A/3B, 9000A/2B This system is designed to acquire CT images, PET images, and fusion images (superimposed PET and CT images). The images that can be acquired include planar images (for any position and from any direction) and volume images (3D images) of the whole human body (including the head). The CT images provide morphological information showing the positional relationships between the lesion and adjacent tissues, while the PET images provide functional information based on the uptake of the injected radioisotope.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Potential degradation of detector performance.
Root cause tag
design
FDA root cause
Device Design
Action
Toshiba sent an Urgent Medical Device Correction letter dated April 24, 2016 to all affected customers. The letter informed customers that Toshiba America Medical Systems is bringing attention to a potential problem with the Celestion PET/CT System. The letter informed the customers of a potential degradation of detector performance and informed the customers of the problems and the corrective actions to be taken. Customers were instructed to complete and return the attached form and fax it to (877) 349-3054 or be sent via email to [email protected]. Customers with questions were instructed to call (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
Quantity in commerce
3 units
Distribution
US Distribution to WI and NV.
Product codes and lots
Serial No. 3BA1572002, 3BA1562001, 2881492002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1607-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1607-2016Class IITerminatedVoluntary: Firm initiated