Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V...
Toshiba American Medical Systems Inc began a recall of Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System on . FDA classified it as Class II and gives the reason as "During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2188-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Toshiba America Medical Systems Angio WorkStation: XIDF-AWS801v6.00, v6.01, and V6.10 System: INFX-8000V Bi-Plane System
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.
- Root cause tag
- software
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Toshiba Medical Systems (TAMS) planned action to bring product into Compliance: 1.TAMS will install the software and perform testing to ensure the software update was effective. 2. The customer notification letter which includes a statement that TAMS will, without charge, remedy the defect or bring the product into compliance. . If you have any questions regarding this, please feel free to contact your local Toshiba representative at (800) 521-1968.
- Quantity in commerce
- 23
- Distribution
- US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.
- Product codes and lots
- Model Number - XIDF-AWS801
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2188-2016 | Class II | Terminated | FDA Mandated |