AB SCIEX Triple Quad 4500MD LC/MS/MS System. Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part...
Ab Sciex began a recall of AB SCIEX Triple Quad 4500MD LC/MS/MS System. Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031257 3200MD series and 4500MD... on . FDA classified it as Class II and gives the reason as "MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2528-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AB Sciex, LLC
- Product
- AB SCIEX Triple Quad 4500MD LC/MS/MS System. Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031257 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.
- Product code
- DOP Mass Spectrometer, Clinical Use
- Reason for recall
- MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- AB SCIEX issued recall notification by letter dated June 16, 2016, to all affected customers. The letter identified the problem, provided temporary measures until a permanent software is available. A response form was to be completed and returned. questions regarding this notice contact SCI EX at +1 289 982 2712.
- Quantity in commerce
- 121 units
- Distribution
- Worldwide Distribution - US (Nationwide) Internationally to Canada, Austria, China, Netherlands, Norway, India, Korea, Germany, France, Switzerland, Ireland, Italy, Singapore, Russia, UK, Malaysia,
- Product codes and lots
- Software Version: MultiQuant MD 3.0, MultiQuant MD 3.0.1 and MultiQuant MD 3.0.2 software
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2526-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2527-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2528-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2529-2016 | Class II | Terminated | Voluntary: Firm initiated |