AB Sciex, LLC

Concord Ontario, CA

AB SCIEX LLC appears in FDA device records in Concord Ontario, CA. FDA records list one 510(k) clearance, no PMA records and 3 registered establishments. As of , FDA records show 12 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
1
PMA records
0
Recall records
12
Registered establishments
3

510(k) clearance history

FDA records hold 1 510(k) clearance for AB Sciex, LLC in 2017.

Clearances by year, as a table
510(k) clearance decisions for AB Sciex, LLC, by decision year
YearClearances
20171
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
DEN170019Vitamin D 200M Assay for the Topaz SystemPSLUnknown

Registered establishments

FDA registered establishments for AB Sciex, LLC, with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
30089355453008935545AB SCIEX LLC250 Forest St, Marlborough, MA 01752 US2026No device listings on this registration
30083368633008336863AB SCIEX LP71 Four Valley Drive, Concord Ontario L4K 4V8 CA2026
30097150513009715051AB SCIEX Pte. Ltd.Blk 33, #04-06, Woodlands Central Industrial Estate Central Singapore 739256 SG2026

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-0040-2024Citrine Triple Quad Mass Spectrometer, Part Number 5063684In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.Class IIOpen, Classified
Z-0039-2024Citrine QTRAP Mass Spectrometer, Part Number 5063685In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.Class IIOpen, Classified
Z-0038-20244500MD Triple Quad Mass Spectrometer, Part Number 5032522In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.Class IIOpen, Classified
Z-0037-2024API3200MD Mass Spectrometer, Part Number 4466230In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.Class IIOpen, Classified
Z-0175-2022Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results.Class IITerminated
Z-2529-2016AB SCIEX QTRAP 4500MD LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.Class IITerminated
Z-2528-2016AB SCIEX Triple Quad 4500MD LC/MS/MS System. Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031257 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.Class IITerminated
Z-2527-2016AB SCIEX 3200MD QTRAP LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5024500 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.Class IITerminated
Z-2526-2016AB Sciex API 3200MD" LC/MS/MS System. In-Vitro Diagnostic Instrument Part Number (REF): 5024501 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.Class IITerminated
Z-1671-2016AB Sciex Triple Quad 4500MD LC/MS/MS System, Part No. 5031257; and QTRAP 4500MD LC/MS/MS System, Part No. 5031231. Mass spectrometers for in vitro diagnostic use only.Under certain conditions, mass spectrometers may report incorrect quantitative results.Class IITerminated
Z-1586-2016Analyst MD Version 1.6.1 and 1.6.2 Software used with the following instruments: API 3200MD" LC/MS/MS System, Instrument Part Number: 5024501; 3200MD QTRAP LC/MS/MS System, Instrument Part Number: 5024500; Triple Quad" 4500MD LC/MS/MS System, Instrument Part Number: 5031257; QTRAP 4500MD LC/MS/MS System, Instrument Part Number: 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.Wrong quantitative results may be displayed in a report from the device, which may potentially lead to an incorrect patient diagnosis.Class IITerminated
Z-2609-2014AB SCIEX API 3200MD" LC/MS/MS System with software: MultiQuantMD 3.0In vitro diagnostic to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens Part Number 5024543.Software- AB SCIEX API 3200MD LC/MS/MS SYSTEM with MultiQuantMD 3.0 software may provide incorrect data for patient diagnosisClass IITerminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets AB Sciex, LLC edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain