510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1397-2017Class II

Toshiba Kalare DREX-KL80 X-ray generator

Canon Medical Systems USA, Inc. / initiated 2016-05-19 / Terminated

Toshiba American Medical Systems Inc began a recall of Toshiba Kalare DREX-KL80 X-ray generator on . FDA classified it as Class II and gives the reason as "It was discovered that the generator of the system could possibly terminate the exposure prematurely during an examination". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1397-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba Kalare DREX-KL80 X-ray generator
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
It was discovered that the generator of the system could possibly terminate the exposure prematurely during an examination
FDA root cause
Radiation Control for Health and Safety Act
Action
Toshiba America Medical Systems planned action to bring defect into compliance : 1. A notification letter will be sent to customers with affected systems. This letter includes the statement that the repairs will be made at no cost to the customer. 2. Modified firmware will be installed on the systems to prevent the defect from recurring. 3. The consignees will be contacted by a Toshiba service representative for an appointment when the modified firmware becomes available 4. The consignee will be informed that if this problem occurs before the new firmware is installed; they should reset the x-ray generator to complete the exam; and 5. The CAP will be completed by May 1, 2017. For further questions please call (714) 730-5000
Quantity in commerce
254
Distribution
USA ( nationwide ) Distribution to the states of : WA, OR, CA, OH, IL, CT, IA, KY, AL, LA, HI, PA, NJ, DE, MN, DC, NY, MI, SC, MD, TN, OK, WI, UT, ID, NV, TX, AL, IN, WV, KS, MO, CO, MT, MS, NC, GA, NH,UT, FL, AZ and ND.
Product codes and lots
Model : KXO-80XD

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1397-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1397-2017Class IITerminatedFDA Mandated
Z-1398-2017Class IITerminatedFDA Mandated