Z-2109-2017Class II

Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F

Canon Medical Systems USA, Inc. / initiated 2017-02-01 / Terminated

Toshiba American Medical Systems Inc began a recall of Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F on . FDA classified it as Class II and gives the reason as "It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2109-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on.
FDA root cause
Radiation Control for Health and Safety Act
Action
Toshiba Medical planned action to be taken to address the defect: CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Any future customer notification letters to be sent must comply with the requirements of 21 CFR 1003.21. This includes, but is not limited to, the prescribed packaging and the statement that you will remedy the defect or bring the product into compliance at no charge. 3. The corrections are made free of charge to customers. 4. You complete this recall by September 30, 2017. Report any failures regarding this recall immediately to the District Recall Coordinator. For further questions, please call (800) 521-1968
Quantity in commerce
US - 70 systems
Distribution
US Distribution to the states of :LA, IL, NY, FL, MA, TN, MO, GA, CA, DE, MO, AL, ID, MN, TX, CT, PA, MS, OH, NC, NV, W VA, WA, AR, NJ, MD, OK, IN and NH.
Product codes and lots
S/N : 8181622393

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2109-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2109-2017Class IITerminatedFDA Mandated