XARIO 100 TUS-X100 diagnostic ultrasound system
Toshiba American Medical Systems Inc began a recall of XARIO 100 TUS-X100 diagnostic ultrasound system on . FDA classified it as Class II and gives the reason as "When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2708-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- XARIO 100 TUS-X100 diagnostic ultrasound system
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.
- Root cause tag
- software
- FDA root cause
- Software Manufacturing/Software Deployment
- Action
- Toshiba Medical sent an Urgent Medical Device Correction letter on May 1, 2017. The consignee is instructed not to use the Continuous Trace method for cardiac velocity trace measurements to quantify the MPG values. Modified software is being provided. A Toshiba service representative will contact the consignee to make an appointment to install the modified software to fix the problem. Customers were asked to complete the form and fax it to the toll-free number shown at the top of the form. The form can also be sent by email to [email protected]. Customers with questions were instructed to call 800-521-1968.
- Quantity in commerce
- 7 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Serial Numbers: W5B16Z2555, W5B1712642, MXA1723100, MXA1723098, MXA1723101, MXA1723099, MXA1723097
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2708-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2709-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2710-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2711-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2712-2017 | Class II | Terminated | Voluntary: Firm initiated |