Z-2710-2017Class II

APLIO 300 TUS-A300 diagnostic ultrasound system

Canon Medical Systems USA, Inc. / initiated 2017-05-01 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of APLIO 300 TUS-A300 diagnostic ultrasound system on . FDA classified it as Class II and gives the reason as "When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2710-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
APLIO 300 TUS-A300 diagnostic ultrasound system
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
When Continuous Trace method is used as the tracing method on Spectral Doppler in velocity trace measurement, an incorrect value may be displayed in the ultrasound systems and in the ultrasound workstations listed below.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Toshiba Medical sent an Urgent Medical Device Correction letter on May 1, 2017. The consignee is instructed not to use the Continuous Trace method for cardiac velocity trace measurements to quantify the MPG values. Modified software is being provided. A Toshiba service representative will contact the consignee to make an appointment to install the modified software to fix the problem. Customers were asked to complete the form and fax it to the toll-free number shown at the top of the form. The form can also be sent by email to [email protected]. Customers with questions were instructed to call 800-521-1968.
Quantity in commerce
90 units
Distribution
Nationwide Distribution
Product codes and lots
Serial Numbers: AAC1692086, AAC1692087, AAC1692088, AAC1692089, AAC1692090, AAC16Y2091, AAC16Y2092, AAC16Y2093, AAC16Y2094, AAC16Y2095, AAC16Y2096, AAC16Y2097, AAC16Y2098, AAC16Y2099, AAC16Y2100, AAD16Z2101, AAD16Z2102, AAD16Z2103, AAD16Z2104, AAD16Z2105, AAD1712106, AAD1712107, AAD1712108, AAD1712109, AAD1712110, AAD1712111, AAD1712112, AAD1722114, AAD1722115, AAD1722116, AAD1722117, AAA1652011, AAA1652012, AAA1652013, AAA1652014, AAA1652015, AAA1672026, AAA1672027, AAA1672028, AAA1672029, AAA1672030, AAA1672031, AAA1672032, AAA1672033, AAA1672034, AAA1672035, AAA1672036, AAA1672037, AAA1672038, AAA1672040, AAA1682041, AAA1682042, AAA1682043, AAA1682044, AAA1682045, AAA1682046, AAA1682047, AAA1682048, AAA1682049, AAA1682051, AAA1682052, AAA1682053, AAA1682054, AAA1692056, AAA1692057, AAA1692058, AAA1692059, AAA1692060, AAA1712011, AAA1712012, AAA1712013, AAA1712014, AAA1712015, AAA1712016, AAA1712017, AAA1722022, AAA1722023, AAA1722024, AAA1722036, AAA1722038, AAA1722039, AAA1722040, AAA1722041, AAA1722042, AAA1722043, AAA1722044, AAA1722045, AAA1732046, AAA1732047, AAA1732049.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2710-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2708-2017Class IITerminatedVoluntary: Firm initiated
Z-2709-2017Class IITerminatedVoluntary: Firm initiated
Z-2710-2017Class IITerminatedVoluntary: Firm initiated
Z-2711-2017Class IITerminatedVoluntary: Firm initiated
Z-2712-2017Class IITerminatedVoluntary: Firm initiated