Z-3070-2017Class II

Toshiba America Medical Systems Ultimax Digital Fluoroscopic X-Ray System Model: DREX-UI80 Ultimax is...

Canon Medical Systems USA, Inc. / initiated 2017-03-15 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of Toshiba America Medical Systems Ultimax Digital Fluoroscopic X-Ray System Model: DREX-UI80 Ultimax is intended for gastrointestinal studies, vascular studies... on . FDA classified it as Class II and gives the reason as "The issue was identified as a firmware problem with the generator". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-3070-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba America Medical Systems Ultimax Digital Fluoroscopic X-Ray System Model: DREX-UI80 Ultimax is intended for gastrointestinal studies, vascular studies, general radiography, and fluoroscopy. The system has medical applications ranging from but not limited to: contrast-enhanced studies, support of endoscopic studies, nonvascular contrast-enhanced studies, nonvascular IVR, vascular contrast-enhanced studies support of vascular NR, and general radiography.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
The issue was identified as a firmware problem with the generator.
Root cause tag
software
FDA root cause
Software design
Action
Toshiba America Medical Systems ( TAMS) will bring the defect into Compliance: 1.You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2.You will install modified firmware on the systems to correct the defect and prevent recurrence. 3.You will provide progress reports on the effectiveness of the plan. For further questions, please call (714) 730-5000.
Quantity in commerce
18
Distribution
US Distribution to the states of : LA, AR, IA, TX, IL, MN. MS, NJ, WV, SC, LA, WY, NC and CA.
Product codes and lots
Ser. No. U1A1692606

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3070-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3070-2017Class IITerminatedFDA Mandated