Z-3111-2017Class II

Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with...

Canon Medical Systems USA, Inc. / initiated 2017-03-23 / Terminated / root cause: software

Toshiba American Medical Systems began a recall of Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with the ceiling suspended tube support, high voltage... on . FDA classified it as Class II and gives the reason as "It was discovered during a procedure that when the operator made an exposure on the wireless x-ray detector and the image data was sent to the digital radiography system, and error message was displayed " System Error...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-3111-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction with the ceiling suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck, and limbs. This system is used for image acquisition, image display and trans mission/ output of images to external devices.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
It was discovered during a procedure that when the operator made an exposure on the wireless x-ray detector and the image data was sent to the digital radiography system, and error message was displayed " System Error (2063)" which required a reboot and loss of the image.
Root cause tag
software
FDA root cause
Software design
Action
Toshiba America Medical Systems Inc. planned action to bring defect into compliance: 1.You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2.You will install modified firmware on the systems to correct the defect and prevent recurrence. 3.You will provide progress reports on the effectiveness of the plan. If you have any questions, please feel free to contact your local Toshiba representative at (800) 421-1968.
Quantity in commerce
140
Distribution
USA (nationwide) Distribution
Product codes and lots
Serial No. W8B1572022

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3111-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3111-2017Class IITerminatedFDA Mandated