510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0012-2018Class II

Toshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal...

Canon Medical Systems USA, Inc. / initiated 2017-09-12 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of Toshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal diagnostic imaging system for radiographic and... on . FDA classified it as Class II and gives the reason as "During an examination images were displayed on the live monitor,but the images were not displayed on the system monitor nor were they saved to the hard disk". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0012-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba Medical Kalare Fluoroscopic X-Ray System (DREX-KL80) Kalare is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. It is intended for use by qualified/trained doctors or technologists on both adult and pediatric subjects taking radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing or lying in prone or supine positions.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
During an examination images were displayed on the live monitor,but the images were not displayed on the system monitor nor were they saved to the hard disk.
Root cause tag
software
FDA root cause
Software design
Action
Toshiba planned action to bring defect into compliance: 1. You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2. You will install modified firmware on the systems to correct the defect and prevent recurrence. 3. You will provide progress reports on the effectiveness of the plan. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator are to be included in the notification. Thank you for your urgent attention to this matter. If you have any questions regarding this letter please feel free to call (800) 421-1968, or contact your local Toshiba Representative at (800) 521-1968. For further questions please call (714) 730-5000.
Quantity in commerce
US - 96
Distribution
USA (nationwide) Distribution
Product codes and lots
Model: HDR-08A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0012-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0012-2018Class IITerminatedFDA Mandated