Z-2315-2018Class II

Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be...

Canon Medical Systems USA, Inc. / initiated 2018-01-04 / Terminated / root cause: software

Toshiba American Medical Systems Inc began a recall of Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is... on . FDA classified it as Class II and gives the reason as "The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2315-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Root cause tag
software
FDA root cause
Software design
Action
Canon sent an Urgent - Medical Device Correction letter dated January 4, 2018. Customers were notified of the issue and informed that a modified firmware will be installed to correct the issue. When that modified firmware becomes available, the Toshiba service representative will contact customers to schedule an appointment to install the firmware. Until the modified firmware is installed, instructions were provided to confirm that the main power switch is sent to OFF, insert the power plug into the power outlet, and to set the main power switch to ON. The system should be operated according to the procedure described in section 9.1, Connection the Power Cable and Protective Grounding, in the operation manual. Additionally, customers were asked to complete and return the attached Customer Reply Form. For further questions, please call (714) 730-5000.
Quantity in commerce
162 units total
Distribution
USA (nationwide) Distribution to the states of : CA, CO, DE, GA, HI, IA, ID, IL, IN, MD, MI, MN, MO, NC, ND, NV, OH, OR, PA, SC, SD, TN, TX, VA, WA, WV and WY.
Product codes and lots
Model No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1732013, HAB1732016, HAB16Z2002, NJC17X2086, NJC17X2087, JHA16X2004, HAB1732027, HAB16Z2001, HAB16Z2002, NJB1762003, JHA1752022, HAB16Z2003, HAB16Z2004, JHA1752019, NJB1782050, HAB1742033, NJB1772031, HNA1792013, HNA1772008, JHA1752010, HAB1742035, HAB1742036, NJB1762001, NJB1782042, JHA1762034, HAB1732030, HAB16Z2005, HAB1732007, HAB1742034, HAB1742037, JHA1752024, NJB1772019, JHA1762029, JHA1762031, NJB1782061, HAB1732019, HNA1772001, HAB1712008, HAB1722011, HAB1712010, HAB1712009, HNA1792014, HAB1722012, HAB16Z2003, HAB1732017, HAB1732018, HAB1712006, HAB1732014, HAB1712007, HAB1732015, HAB1732021, HAB1732020, HAB1732025, JHA1762030, HAB1732029, HAB1732031, NJC17X2066, HNA1772002, NJB1782060, NJB1782043, NJB1762002, JHA16X2002, NJB1782057, NJB1782055, JHA1752021, NJB1782058, NJB1762006, NJB1782056, NJC17X2089, NJB1772023, NJB1782051, NJB1782052, NJB1782053, NJB1782054, HAB1732006, NJB1782062, NJB1782063, NJB1762004, NJB1762005, NJB1772022, JHA1752014, JHA1752016, JHA1752017, JHA1752018, HNA1792012, NJC17X2068, NJC17X2069, NJC17X2070, NJC17X2067, NJB1782046, NJB1782047, NJB1782048, NJB1782049, NJB1772039, NJB1782041, NJB1782045, NJC17X2083, NJC17X2084, NJC17X2081, NJC17X2079, NJC17X2080, NJC17X2082, NJC17X2076, NJC17X2077, NJC17X2078, NJB1772021, NJB1782044, JHA1752020, HNA1772005, NJB1782064, JHA1762032, HAB1732010, HAB1732008, NJB1782059, NJB1762008, NJB1762007, NJB1762009, NJB1772024, NJB1772014, NJB1772017, NJB1772018, NJB1772015, NJB1772016, NJB1772011, NJB1772012, NJB1772020, NJB1762010, NJB1772035, NJB1772036, NJB1772037, NJB1772038, NJB1772026, NJB1772027, NJB1772028, NJB1772030, NJB1772034, NJB1772025, NJB1772032, NJB1772013, JHA1762036, NJC17X2090, JHA1762033, HNA1772010, HNA1772009, NJB1772029, HNA1772004, HNA1772003, NJB1782065, NJC17X2071, NJB17822040, NJC17X2074, NJC17X2075, NJC17X2085, HAB1732026, JHA16X2007, HAB1732022, JHA1752023, JHA1762026, JHA1762025, HAB1732028, JHA16X2001, JHA16X2003, HAB1742011, HAB1742012, JHA16X2001, JHA16X2003, JHA1752004, JHA1752005, JHA1752006, JHA1752007, JHA1752008, JHA1752009, JHA1752010, JHA1752011, JHA1752012, JHA1752013, NJB1772033, NJC17X2072, NJC17X2073, NJC17X2088, HAB16Z2001, HAB16Z2004, HAB16Z2005, HAB1732024, HAB1732032, HNA1792015, HNB17X2016, HNB17X2017, HNB17X2018, HNB17X2019, HNB17X2020, HNB17X2021, HNB17X2022, HNB17X2023, HNB17X2024, HNB17X2025, HNB17X2026, HNB17X2027, HNB17X2028, HNB17X2029, HNB17X2030, JHA16X2002, JHA16X2005, JHA16X2006, JHA16X2008, JHA16X2009, JHA1752011, JHA1752012, JHA1752013, JHA1752015, JHA1762027, JHA1762028, JHA1762035, JHA1762037, JHA1762038, JHA1762039, HAB1742001, HAB1742002, JHA16X2001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2315-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2315-2018Class IITerminatedVoluntary: Firm initiated