Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be...
Toshiba American Medical Systems Inc began a recall of Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is... on . FDA classified it as Class II and gives the reason as "The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2315-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Canon sent an Urgent - Medical Device Correction letter dated January 4, 2018. Customers were notified of the issue and informed that a modified firmware will be installed to correct the issue. When that modified firmware becomes available, the Toshiba service representative will contact customers to schedule an appointment to install the firmware. Until the modified firmware is installed, instructions were provided to confirm that the main power switch is sent to OFF, insert the power plug into the power outlet, and to set the main power switch to ON. The system should be operated according to the procedure described in section 9.1, Connection the Power Cable and Protective Grounding, in the operation manual. Additionally, customers were asked to complete and return the attached Customer Reply Form. For further questions, please call (714) 730-5000.
- Quantity in commerce
- 162 units total
- Distribution
- USA (nationwide) Distribution to the states of : CA, CO, DE, GA, HI, IA, ID, IL, IN, MD, MI, MN, MO, NC, ND, NV, OH, OR, PA, SC, SD, TN, TX, VA, WA, WV and WY.
- Product codes and lots
- Model No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1732013, HAB1732016, HAB16Z2002, NJC17X2086, NJC17X2087, JHA16X2004, HAB1732027, HAB16Z2001, HAB16Z2002, NJB1762003, JHA1752022, HAB16Z2003, HAB16Z2004, JHA1752019, NJB1782050, HAB1742033, NJB1772031, HNA1792013, HNA1772008, JHA1752010, HAB1742035, HAB1742036, NJB1762001, NJB1782042, JHA1762034, HAB1732030, HAB16Z2005, HAB1732007, HAB1742034, HAB1742037, JHA1752024, NJB1772019, JHA1762029, JHA1762031, NJB1782061, HAB1732019, HNA1772001, HAB1712008, HAB1722011, HAB1712010, HAB1712009, HNA1792014, HAB1722012, HAB16Z2003, HAB1732017, HAB1732018, HAB1712006, HAB1732014, HAB1712007, HAB1732015, HAB1732021, HAB1732020, HAB1732025, JHA1762030, HAB1732029, HAB1732031, NJC17X2066, HNA1772002, NJB1782060, NJB1782043, NJB1762002, JHA16X2002, NJB1782057, NJB1782055, JHA1752021, NJB1782058, NJB1762006, NJB1782056, NJC17X2089, NJB1772023, NJB1782051, NJB1782052, NJB1782053, NJB1782054, HAB1732006, NJB1782062, NJB1782063, NJB1762004, NJB1762005, NJB1772022, JHA1752014, JHA1752016, JHA1752017, JHA1752018, HNA1792012, NJC17X2068, NJC17X2069, NJC17X2070, NJC17X2067, NJB1782046, NJB1782047, NJB1782048, NJB1782049, NJB1772039, NJB1782041, NJB1782045, NJC17X2083, NJC17X2084, NJC17X2081, NJC17X2079, NJC17X2080, NJC17X2082, NJC17X2076, NJC17X2077, NJC17X2078, NJB1772021, NJB1782044, JHA1752020, HNA1772005, NJB1782064, JHA1762032, HAB1732010, HAB1732008, NJB1782059, NJB1762008, NJB1762007, NJB1762009, NJB1772024, NJB1772014, NJB1772017, NJB1772018, NJB1772015, NJB1772016, NJB1772011, NJB1772012, NJB1772020, NJB1762010, NJB1772035, NJB1772036, NJB1772037, NJB1772038, NJB1772026, NJB1772027, NJB1772028, NJB1772030, NJB1772034, NJB1772025, NJB1772032, NJB1772013, JHA1762036, NJC17X2090, JHA1762033, HNA1772010, HNA1772009, NJB1772029, HNA1772004, HNA1772003, NJB1782065, NJC17X2071, NJB17822040, NJC17X2074, NJC17X2075, NJC17X2085, HAB1732026, JHA16X2007, HAB1732022, JHA1752023, JHA1762026, JHA1762025, HAB1732028, JHA16X2001, JHA16X2003, HAB1742011, HAB1742012, JHA16X2001, JHA16X2003, JHA1752004, JHA1752005, JHA1752006, JHA1752007, JHA1752008, JHA1752009, JHA1752010, JHA1752011, JHA1752012, JHA1752013, NJB1772033, NJC17X2072, NJC17X2073, NJC17X2088, HAB16Z2001, HAB16Z2004, HAB16Z2005, HAB1732024, HAB1732032, HNA1792015, HNB17X2016, HNB17X2017, HNB17X2018, HNB17X2019, HNB17X2020, HNB17X2021, HNB17X2022, HNB17X2023, HNB17X2024, HNB17X2025, HNB17X2026, HNB17X2027, HNB17X2028, HNB17X2029, HNB17X2030, JHA16X2002, JHA16X2005, JHA16X2006, JHA16X2008, JHA16X2009, JHA1752011, JHA1752012, JHA1752013, JHA1752015, JHA1762027, JHA1762028, JHA1762035, JHA1762037, JHA1762038, JHA1762039, HAB1742001, HAB1742002, JHA16X2001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2315-2018 | Class II | Terminated | Voluntary: Firm initiated |