Z-2077-2019Class II

PulsioFlex Monitoring System, Part Number: 6882747, Model Number: PC400 3 R0. Usage: The PulsioFlex...

Getinge US Sales LLC / initiated 2019-06-05 / Terminated / root cause: software

GETINGE US SALES LLC began a recall of PulsioFlex Monitoring System, Part Number: 6882747, Model Number: PC400 3 R0. Usage: The PulsioFlex Monitoring System is indicated in patients where... on . FDA classified it as Class II and gives the reason as "The monitor displays an error message . The error message states "internal error restart or service"". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2077-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Getinge US Sales LLC
Product
PulsioFlex Monitoring System, Part Number: 6882747, Model Number: PC400 3 R0. Usage: The PulsioFlex Monitoring System is indicated in patients where cardiovascular and circulatory volume status monitoring is necessary. If a patient s biometric data are entered, the PulsioFlex Monitor presents the derived parameters indexed. -With the PiCCO Module cardiac output is determined both continuously through pulse contour analysis and intermittently through thermodilution technique. Both are used for the determination of other derived parameters. -With the CeVOX oximetry module connected to a compatible oximetry probe, the PulsioFlex Monitoring System measures continuous venous oxygen saturation to assess oxygen delivery and consumption. The use of the PulsioFlex Monitoring System is indicated in patients where cardiovascular and organ monitoring is useful. This includes patients in surgical, medical, and other hospital units
Product code
DXG Computer, Diagnostic, Pre-Programmed, Single-Function
Reason for recall
The monitor displays an error message . The error message states "internal error restart or service".
Root cause tag
software
FDA root cause
Software design
Action
PULSION Medical Systems SE /Getinge issued a customer notification (i.e. Urgent Medical Device correction/ Removal) to the affected US consignees via FedEx Priority Overnight Delivery with Signature Proof of Delivery. 1. Examine your inventory immediately to determine if you have any affected PulsioFlex Monitors. 2.Should you have one or more PulsioFlex Monitors you will need to return the device(s) for service. 3. contact PULSION Medical Systems SE/Getinge Sales Support at 1-888-627-8383 (press option 2, then option 2) to request a return authorization (RMA) and shipping instructions to return any affected product. 4. Please complete and sign the attached MEDICAL DEVICE RECALL (REMOVAL) - RESPONSE FORM (page 4) to acknowledge that you have received this notification. 5. Return the completed form to PULSION Medical Systems SE/Getinge Sales Support by e-mailing a scanned copy to [email protected] or by faxing the form to 1-800-862-5307. If you have any questions, please contact PULSION Medical Systems SE/Getinge Sales Support at 1-888-627-8383 (press option 2, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone)
Quantity in commerce
7
Distribution
US Nationwide distribution in the states of MD, OK, NY,FL, SC
Product codes and lots
software version:V5.1.0.7 US ver. A.,   Serial number affected: K17400011005 , K17400011065, K17400011082, K17400011086, K17400011097, K17400011107, L17400011267.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2077-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2077-2019Class IITerminatedVoluntary: Firm initiated