US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14...
Flowonix Medical Inc began a recall of US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082 on . FDA classified it as Class II and gives the reason as "A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance"...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1123-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Flowonix Medical Inc
- Product
- US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- ***Updated 5/12/2023*** On May 11, 2023, the firm sent an URGENT MEDICAL DEVICE CORRECTION UPDATE letter to all affected customers, which informed them of an update to Flowonix's plan to correct the firmware issue. Flowonix will NOT be conducting an office-based pump firmware update for affected pumps implanted prior to January 2020. Healthcare providers were provided ongoing recommendations for Prometra and Prometra II pumps implanted prior to January 2020, which are impacted by this issue. Healthcare providers should contact their Flowonix Representative or our Technical Solutions Department (1-855-356-9665) in the event cessation of therapy or premature low battery is suspected and/or have general questions regarding this notice. *** On December 24, 2019, the firm notified affected users via an Urgent Medical Device Corrective Action Recall Notice. The firm advised customers of the product issue and reminded customers that the pumps are currently labeled only for use with Infumorph. Due to the low risk that this failure mode may occur, Flowonix is not recommending that pumps be explanted. Healthcare providers were recommended to inform patients regarding the symptoms and adverse health events associated with an unexpected pump stoppage and the appropriate steps to take in the event they suspect a stoppage. The firm will be implementing a correction of firmware errors to bring the pumps back into specification. The firm will follow up when the change has been completed. An office-based pump firmware update is being developed for patients currently implanted with the affected pumps. (Note, as of the 5/12/23 update, implanted pumps will not be receiving a firmware update.)
- Quantity in commerce
- 6670 (US)
- Distribution
- Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.
- Product codes and lots
- All units with pump firmware version 0.26 manufactured from 2012 until the present time. Serial Numbers 10HW4A57 10AX4A06 36ES4A15 36AR4A04 36AR4A06 36AR4A13 36HS4A10 36HS4A54 36JN4A04 36JN4A06 36JN4A07 36JN4A09 36JN4A13 36JN4A14 36JN4A15 36JN4A18 36JN4A23 36JN4A28 36JN4A35 36KN4A11 36KN4A16 36KN4A23 36KN4A27 36KN4A33 36KN4A34 36KN4A41 36HS4A01 10IT4A21 36HS4A14 36HS4A26 36BS4A17 36ES4A28 36ES4A33 36ES4A08 10IT4A16 36FR4A13 36HS4A61 10ET4A07 36HS4A46 10ET4A20 36HS4A25 36HS4A43 10ET4A02
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1120-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1121-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1122-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1123-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1124-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1125-2020 | Class II | Ongoing | Voluntary: Firm initiated |