Flowonix Medical Inc
Mansfield, MA, US
Flowonix Medical Inc appears in FDA device records in Mansfield, MA. FDA records list no 510(k) clearances, no PMA records and no registered establishments. As of , FDA records show 13 recalls for this company.
FDA records held for this company
- 510(k) clearances
- 0
- PMA records
- 0
- Recall records
- 13
- Registered establishments
- 0
510(k) clearance history
No 510(k) clearance records for this company in the ingested FDA data.
No 510(k) clearance records in the ingested FDA data.
Registered establishments
No FDA establishment registrations recorded against this company.
FDA product codes
No FDA product codes recorded against this company.
Recall records
| Recall | Product | Reason | Classification | Initiated | Status |
|---|---|---|---|---|---|
| Z-1996-2021 | Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242 | Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection | Class II | Terminated | |
| Z-1125-2020 | Prometra II Programmable 20mL Pump. Catalog No. 93827 The device is an implantable infusion pump. GTIN-14 00810335020YYY (YYY= in-country language specific) | A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system. | Class II | Open, Classified | |
| Z-1124-2020 | Prometra Programmable 20mL Pump. Catalog No. 91827 The device is an implantable infusion pump. GTIN-14 00810335020XXX (XXX = in-country language specific) | A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system. | Class II | Open, Classified | |
| Z-1123-2020 | US Clinical 20mL Programmable Pump. Catalog No. 01827 The device is an implantable infusion pump. GTIN-14 00810335020082 | A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system. | Class II | Open, Classified | |
| Z-1122-2020 | Prometra II 40mL Pump. Catalog No. 16827 The device is an implantable infusion pump. GTIN-14 00810335020242 | A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system. | Class II | Open, Classified | |
| Z-1121-2020 | Prometra II 20mL Pump. Catalog No. 13827 The device is an implantable infusion pump. GTIN-14 00810335020228 | A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system. | Class II | Open, Classified | |
| Z-1120-2020 | Prometra Programmable 20mL Pump. Catalog No. 11827 The device is an implantable infusion pump. GTIN-14 00810335020082 | A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system. | Class II | Open, Classified | |
| Z-1716-2019 | Prometra II Programmable Pump (Clinical), Cat. No. 01827 For intrathecal infusion. | The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped. | Class II | Terminated | |
| Z-1715-2019 | Prometra II Programmable Pump, Cat. No. 13827 For intrathecal infusion. | The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped. | Class II | Terminated | |
| Z-1714-2019 | Prometra Programmable Pump, Cat. No. 11827 For intrathecal infusion. | The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped. | Class II | Terminated | |
| Z-1713-2019 | Prometra Programmer (Grand Prime) Software version 2.00.29, Cat. No. 13828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion. | The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. | Class II | Terminated | |
| Z-1712-2019 | Prometra Clinician Programmer (Touchscreen) Software version 2.00.29, Cat. No. 12828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion. | The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. | Class II | Terminated | |
| Z-2740-2017 | PROMETRA II Programmable Infusion Pump, Implantable Infusion Pump, Catalog Number/REF 13827, Part Number PL-31790-02. Manufactured by: Flowonix Medical Inc., --- The correction involves the labeling of the device plus the following Prometra II pump labeling is also being revised: Intrathecal Catheter IFU, REF 11823, PL-22790-02; Catheter Revision Kit IFU, REF 11 830, PL-21798-02; Refill Kit IFU, REF 11825, PL-21794-00; Prometra II Patient Implant Card (Permanent) PL-32300-01; Prometra II Patient Implant Card (Temporary): PL-32375-01; Prometra II Patient Guide: PL-31912-01. The Prometra Programmable Infusion Pump System is indicated for intrathecal infusion of Infumorph (preservative-free morphine sulfate sterile solution) or preservative-free sterile 0.9% saline solution (sodium chloride injection, USP). | Flowonix Medical received a report of a patient implanted with the Prometra II Programmable Pump who may have received a fatal drug overdose during an MRI procedure. The Prometra II pump has an FDA-approved design feature intended to permit safe exposure to an MRI without removing drug from the reservoir. | Class II | Terminated |
Verification record
There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.
Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.
MDM coverage
MDM coverage of this company lists here, newest first. Submit news.
Is this your company?
This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.
Claiming lets Flowonix Medical Inc edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.
Data as of , FDA export datePage updated Source: openFDA, public domain