Medtronic Simulus 750 Flexible Ring/Band Accessory Kit -
Medtronic CoreValve LLC began a recall of Medtronic Simulus 750 Flexible Ring/Band Accessory Kit - Product Usage: Used to size a patient's mitral or tricuspid valve in cardiac procedures on . FDA classified it as Class II and gives the reason as "Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1449-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Heart Valves Division
- Product
- Medtronic Simulus 750 Flexible Ring/Band Accessory Kit - Product Usage: Used to size a patient's mitral or tricuspid valve in cardiac procedures.
- Product code
- DTI Sizer, Heart-Valve, Prosthesis
- Reason for recall
- Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm initiated the recall by email or verbal contact on 01/14/2020. The direct consignees (sales representatives) were directed to quarantine and return the affected product. On 02/05/2020, consignees (hospitals) were notified by letter delivered by the Medtronic Sales Representatives. The notices directed the consignee to not use the affected product for clinical use, quarantine and return affected product. Affected product will be replaced by Medtronic.
- Quantity in commerce
- 6
- Distribution
- US Nationwide distribution in the states of Ohio and Kansas.
- Product codes and lots
- GTIN: 00643169707535
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1449-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1450-2020 | Class II | Ongoing | Voluntary: Firm initiated |