Z-1682-2020Class II

McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management

Change Healthcare Canada Company / initiated 2020-02-10 / Open, Classified / root cause: design

CHANGE HEALTHCARE CANADA COMPANY began a recall of McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management on . FDA classified it as Class II and gives the reason as "Inaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG format". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1682-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Change Healthcare Canada Company
Product
McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Inaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG format
Root cause tag
design
FDA root cause
Device Design
Action
Change Healthcare Canada Company notified customers on about 02/10/2020 via "Urgent Field Safety Notice" letter. Customers were informed of an issue whereby users who are estimating heart rate by counting grid squares, may assume that the Sequential ECG format is presented on a scale of 25 mm/sec and thereby may inaccurately estimate the heart rate value. Instructions included that the Simultaneous ECG format can be used rather than the Sequential ECG format, until the software is updated and the Sequential ECG format option can be used while maintaining awareness to the scale bar on the screen. Customers were also instructed to contact their users and train them on how to verify the ECG format mode they are working on and how they can change it. The notice needs to be distributed to all those within their organization who need to be aware and to please alert other parties affiliated with your organization that are affected by this field safety notice. Additionally, customers were informed that Change Healthcare is developing a software update that will enable presentation of the Sequential ECG format view using the 25 mm/sec scale and were requested to contact Change Healthcare Customer Support to: 1) Acknowledge they have read and understood this Urgent Field Safety Notice and 2) Arrange installation of software update/upgrade. Change Healthcare Customer Support contacts: - US / Canada: 1 866-777-0202 - UK / Ireland: +44 (0) 208 9527399
Quantity in commerce
51 units
Distribution
Distribution US nationwide, Canada and the United Kingdom
Product codes and lots
McKesson Cardiology Software Versions: 13.2.1 HF1, 13.2.1 HF2, 14.0, 14.0 HF1, 14.0 HF2;   Change Healthcare Software Versions: 14.1.1, 14.2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1682-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1682-2020Class IIOngoingVoluntary: Firm initiated