Z-1682-2020Class II
McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management
Change Healthcare Canada Company / initiated 2020-02-10 / Open, Classified / root cause: design
CHANGE HEALTHCARE CANADA COMPANY began a recall of McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management on . FDA classified it as Class II and gives the reason as "Inaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG format". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1682-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Change Healthcare Canada Company
- Product
- McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Inaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG format
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Change Healthcare Canada Company notified customers on about 02/10/2020 via "Urgent Field Safety Notice" letter. Customers were informed of an issue whereby users who are estimating heart rate by counting grid squares, may assume that the Sequential ECG format is presented on a scale of 25 mm/sec and thereby may inaccurately estimate the heart rate value. Instructions included that the Simultaneous ECG format can be used rather than the Sequential ECG format, until the software is updated and the Sequential ECG format option can be used while maintaining awareness to the scale bar on the screen. Customers were also instructed to contact their users and train them on how to verify the ECG format mode they are working on and how they can change it. The notice needs to be distributed to all those within their organization who need to be aware and to please alert other parties affiliated with your organization that are affected by this field safety notice. Additionally, customers were informed that Change Healthcare is developing a software update that will enable presentation of the Sequential ECG format view using the 25 mm/sec scale and were requested to contact Change Healthcare Customer Support to: 1) Acknowledge they have read and understood this Urgent Field Safety Notice and 2) Arrange installation of software update/upgrade. Change Healthcare Customer Support contacts: - US / Canada: 1 866-777-0202 - UK / Ireland: +44 (0) 208 9527399
- Quantity in commerce
- 51 units
- Distribution
- Distribution US nationwide, Canada and the United Kingdom
- Product codes and lots
- McKesson Cardiology Software Versions: 13.2.1 HF1, 13.2.1 HF2, 14.0, 14.0 HF1, 14.0 HF2; Change Healthcare Software Versions: 14.1.1, 14.2
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1682-2020 | Class II | Ongoing | Voluntary: Firm initiated |