System Model: Ultimax-i, DREX-UI80 (DREX-UI80); Unit Model: HDR-08A Imager Processor; UDI (01)04987670100147
Canon Medical System, USA, INC. began a recall of System Model: Ultimax-i, DREX-UI80 (DREX-UI80); Unit Model: HDR-08A Imager Processor; UDI (01)04987670100147 on . FDA classified it as Class II and gives the reason as "During a procedure, when images were acquired, and these images were transferred to PACS /image archival system, the number of images displayed were not the same as originally acquired". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2885-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- System Model: Ultimax-i, DREX-UI80 (DREX-UI80); Unit Model: HDR-08A Imager Processor; UDI (01)04987670100147
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- During a procedure, when images were acquired, and these images were transferred to PACS /image archival system, the number of images displayed were not the same as originally acquired.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recall letters were sent via facsimile between 07/08-10/2020. The letter contains a return reply form that is to be completed by the customer and faxed to CMSU for retention. After approximately one month contact will be made with each non-responding location and a copy of the letter will be faxed to that location and the fax receipt will serve as proof of notification.
- Quantity in commerce
- 11 units
- Distribution
- US - CA, CT, ID, LA, MT, NV, OH, SC, WI and WV
- Product codes and lots
- The following serial numbers are affected: U4D1992027, U4D19X2029, U4D19Y2030, U4D19Y2031, U4D19Y2033, U4E2012035, U4E2012036, U4E2022037, U4D19Y2032, U4E19Z2034 and U4E2032041.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2885-2020 | Class II | Ongoing | Voluntary: Firm initiated |