Z-0549-2021Class II

Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L -

Canon Medical Systems USA, Inc. / initiated 2020-11-09 / Terminated / root cause: software

Canon Medical System, USA, INC. began a recall of Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body... on . FDA classified it as Class II and gives the reason as "A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0549-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.
Root cause tag
software
FDA root cause
Software design
Action
On 11/9/2020, Canon issued a "Urgent: Medical Device Correction" letter to all affected customers via facsimile. In addition, to informing consignees about the recall, the firm ask affected consignees to take the following actions: 1. Until corrective action has been taken, in order to avoid the problem, please use the system with setting InstaView OFF according to the last notification (dated June 2020) Subject: Abnormality in Realtime Reconstruction Images. Please see Attachment A. 2. Please share this information with all users and reviewing radiologists as well as the clinical engineering or biomedical group at your facility. 3. If you have any questions, please feel free to contact Terry Schultz Regulatory Affairs Manager at (800) 421-1968 or your local Canon service representative at (800) 521-1968.
Quantity in commerce
60 units
Distribution
US Nationwide distribution. OUS: Unknown
Product codes and lots
Serial Numbers: 8LA2052063,  8LA19Z2034 ,  8LA19X2004,  8LA19Z2027,  8LA19Z2029,  8LA2052062,  8LA2042055,  8LA19Z2032,  8OA2052059,  8LA2042051,  8LA2032044,  8LA19X2002,  1LB1922175,  8LA2042050,  8LA19X2007,  8LA2012037,  8LA19Y2021,  8LA19Y2014,  8LA2042056,  8LA19Z2023,  8LA2052067,  8LA19Z2031,  8LA19Y2019,  8LA19Z2030,  8OA19Y2013,  8LA19Z2028,  8LA19Y2012,  8LA19Y2020,  8LA2032042,  8LA2052066,  8LA19X2003,  8LA19Y2022,  8LA19X2008,  8LA2042047,  8LA19Z2024,  8LA2032046,  8LA19Z2033,  8OA2042053,  8LA2012036,  8LA2042054,  8LA19Z2025,  8LA19Z2026,  8LA2022041,  8LA19X2010,  8LA2032043,  8OA19X2001,  8OA2052061,  8OA2042049,  8LA19Y2016,  8LA2022039,  8LA19Y2015,  8LA2012035, 8LA2032045,  8LA19Y2018,  8LA19X2005,  8LA19X2006,  8LA19X2009,  8LA19Y2011,  8LA19Y2017,  8LA2012038,  8LA2052060,  8LA2052058 ,  8LA2052064,  8LA2052065,  8LA2052068,  8LA2022040,  8OA2042052,  8LA2042048,  8LA2052057

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0549-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0549-2021Class IITerminatedVoluntary: Firm initiated