Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L -
Canon Medical System, USA, INC. began a recall of Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body... on . FDA classified it as Class II and gives the reason as "A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0549-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 11/9/2020, Canon issued a "Urgent: Medical Device Correction" letter to all affected customers via facsimile. In addition, to informing consignees about the recall, the firm ask affected consignees to take the following actions: 1. Until corrective action has been taken, in order to avoid the problem, please use the system with setting InstaView OFF according to the last notification (dated June 2020) Subject: Abnormality in Realtime Reconstruction Images. Please see Attachment A. 2. Please share this information with all users and reviewing radiologists as well as the clinical engineering or biomedical group at your facility. 3. If you have any questions, please feel free to contact Terry Schultz Regulatory Affairs Manager at (800) 421-1968 or your local Canon service representative at (800) 521-1968.
- Quantity in commerce
- 60 units
- Distribution
- US Nationwide distribution. OUS: Unknown
- Product codes and lots
- Serial Numbers: 8LA2052063, 8LA19Z2034 , 8LA19X2004, 8LA19Z2027, 8LA19Z2029, 8LA2052062, 8LA2042055, 8LA19Z2032, 8OA2052059, 8LA2042051, 8LA2032044, 8LA19X2002, 1LB1922175, 8LA2042050, 8LA19X2007, 8LA2012037, 8LA19Y2021, 8LA19Y2014, 8LA2042056, 8LA19Z2023, 8LA2052067, 8LA19Z2031, 8LA19Y2019, 8LA19Z2030, 8OA19Y2013, 8LA19Z2028, 8LA19Y2012, 8LA19Y2020, 8LA2032042, 8LA2052066, 8LA19X2003, 8LA19Y2022, 8LA19X2008, 8LA2042047, 8LA19Z2024, 8LA2032046, 8LA19Z2033, 8OA2042053, 8LA2012036, 8LA2042054, 8LA19Z2025, 8LA19Z2026, 8LA2022041, 8LA19X2010, 8LA2032043, 8OA19X2001, 8OA2052061, 8OA2042049, 8LA19Y2016, 8LA2022039, 8LA19Y2015, 8LA2012035, 8LA2032045, 8LA19Y2018, 8LA19X2005, 8LA19X2006, 8LA19X2009, 8LA19Y2011, 8LA19Y2017, 8LA2012038, 8LA2052060, 8LA2052058 , 8LA2052064, 8LA2052065, 8LA2052068, 8LA2022040, 8OA2042052, 8LA2042048, 8LA2052057
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0549-2021 | Class II | Terminated | Voluntary: Firm initiated |