Z-1773-2021Class II

Change Healthcare Enterprise Viewer -

Change Healthcare Canada Company / initiated 2021-04-19 / Open, Classified / root cause: software

CHANGE HEALTHCARE CANADA COMPANY began a recall of Change Healthcare Enterprise Viewer - Product Usage: intended to be used with off-the-shelf hardware for the 2D and 3D display of DICOM and non-DICOM medical... on . FDA classified it as Class II and gives the reason as "A software defect was identified where the Image Styles defined by presentation states from CHRS are not displayed in CHEV". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1773-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Change Healthcare Canada Company
Product
Change Healthcare Enterprise Viewer - Product Usage: intended to be used with off-the-shelf hardware for the 2D and 3D display of DICOM and non-DICOM medical images, reports, and multimedia content.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
A software defect was identified where the Image Styles defined by presentation states from CHRS are not displayed in CHEV.
Root cause tag
software
FDA root cause
Software design
Action
On 04/19/21 the firm sent a letter to its consignees with the following instructions: Immediate actions to take to reduce the risk to patient Prior to treating patients, always confirm the correct patient anatomy by reviewing the laterality markers applied during image acquisition. When the issue is detected, the clinician is able to manually adjust the Image Styles in CHEV to accommodate their viewing requirements and/or orient images to standard anatomical position. Product updates that will address the issue Change Healthcare is developing a software update to correct image display of orientation, image inversion and window level information from Image Styles. Recommended actions to take on potentially affected data We strongly encourage customers receiving this Urgent Field Safety Notice to immediately contact Change Healthcare Customer Support to: 1) Acknowledge they have read and understood this Urgent Field Safety Notice; and, 2) Arrange the installation of software update. Change Healthcare Customer Support: 1-800-663-2533 Menu Option 3 (CHEV)
Quantity in commerce
33 devices
Distribution
US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IA, IN, IL, KY, NE, NY, MA, MI, MS, OH, PA, TX,TN, WA, WI, WY.
Product codes and lots
Software version 2.0 and 2.1.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1773-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1773-2021Class IIOngoingVoluntary: Firm initiated