Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra...
Flowonix Medical Inc began a recall of Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s)... on . FDA classified it as Class II and gives the reason as "Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1996-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Flowonix Medical Inc
- Product
- Physician Order Form (PL-15400-00, Nov. 2018) distributed to end user used with the following: Prometra Programmable Infusion Pump System: MODEL/CATALOG #(s): Catalog No. US Model GTIN-14 11827 Prometra Programmable 20mL Pump 00810335020082 13827 Prometra II Programmable 20mLPump 00810335020228 16827 Prometra II Programmable 40mLPump 00810335020242
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Flowonix Customer letters) have been mailed via USPS 2 day priority mail with electronic tracking on 5/26-5/27. Letter states reason for recall, health risk and Actions to be taken by the Customer: 1) Discontinue the use and distribution of the Physician Order Form (PL-15400-00, Nov. 2018). This form is no longer available for use and shall no longer be used. 2) Complete and return the attached response form to Flowonix. Should you have any questions, please contact your Flowonix Representative or our Technical Solutions Department at (855) 356-9665
- Distribution
- US Nationwide distribution.
- Product codes and lots
- All Physician Order Form (PL-15400-00, Nov. 2018)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1996-2021 | Class II | Terminated | Voluntary: Firm initiated |