Z-2043-2021Class II
CoreValve Evolut PRO PLUS Delivery Catheter System, REF: D-EVPROP2329US; D-EVPROP34US
Medtronic Heart Valves Division / initiated 2021-06-11 / Open, Classified / root cause: design
Medtronic Heart Valves Division began a recall of CoreValve Evolut PRO PLUS Delivery Catheter System, REF: D-EVPROP2329US; D-EVPROP34US on . FDA classified it as Class II and gives the reason as "Due to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2043-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Heart Valves Division
- Product
- CoreValve Evolut PRO PLUS Delivery Catheter System, REF: D-EVPROP2329US; D-EVPROP34US;
- Product code
- NPT Aortic Valve, Prosthesis, Percutaneously Delivered
- Reason for recall
- Due to reports of actuator separation for various delivery catheter systems. Affected products were manufactured before July 11, 2020.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 06/09/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via UPS 2-day (US Customers) and regionally approved methods (OUS Customer) to customers informing them that it has been identified that were is the potential for actuator separation with delivery catheter systems manufactured before 07/11/2020. Customer are instructed to: 1. Identify and quarantine all unused affected product manufactured before 11-JUL-2020.The enclosed Customer Notification Detail Report includes affected product for their facility. 2. Return all unused affected product in their inventory to Recalling Firm. Contact Customer Service at 800-854-3570 to initiate a product return. Their local Representative can assist them as necessary in initiating the return of this product. 3. Complete the enclosed Customer Confirmation Form and email it to [email protected]. 4. Forward this notice to all who need to be aware within their organization and to any organization where the affected product may have been transferred. For any questions or assistance, contact Customer Service at 800-854-3570
- Quantity in commerce
- Total of all Delivery Systems (Globally) = 67,339 units
- Distribution
- U.S.: AL, AK, AR, AZ, CA, CT, DE, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NE, NH, NJ, NM, NV, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Albania, Argentina, Australia, Austria, Belarus, Belgium, Brazil, BULGARIA, Canada, Chile, CHINA, Colombia, COSTA RICA, Croatia, Curacao, CYPRUS, Czech Republic, Denmark, DOMINICAN REPUBLIC, Egypt, ESTONIA, FINLAND, FRANCE, FRENCH POLYNESIA, GEORGIA, Germany, Greece, GUATEMALA, HONDURAS, Hong Kong, Hungary, ICELAND, India, INDONESIA, IRAN (ISLAMIC REPUBLIC OF), Ireland, Israel, Italy, Japan, JORDAN, KAZAKHSTAN, KOREA, REPUBLIC OF, KUWAIT, LATVIA, LEBANON, LITHUANIA, Luxembourg, MACEDONIA (THE FORMER YUGOSLAV REPUBLIC OF), Malaysia, MALTA, Mexico, MOLDOVA (REPUBLIC OF), Montenegro, Netherlands, New Zealand, Norway, OMAN, Panama, Peru, Poland, Portugal, QATAR, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, SLOVAKIA (Slovak Republic), SLOVENIA, South Africa, Spain, Sweden, Switzerland, SYRIAN ARAB REPUBLIC, Taiwan (Republic of China), THAILAND, Turkey, UNITED ARAB EMIRATES, United Kingdom, URUGUAY, UZBEKISTAN, VENEZUELA and Viet Nam.
- Product codes and lots
- Model Number/Description: D-EVPROP2329US/DELIV SYS D-EVPROP2329US GTIN:00763000207854 Lot Numbers: 0010165308; 0010165310; 0010165312; 0010165313; 0010169462; 0010169464; 0010171866; 0010171868; 0010171869; 0010173361; 0010175293; 0010175294; 0010181410; 0010181411; 0010181412; 0010182735; 0010182736; 0010188541; 0010192246; 0010192247; 0010197788; 0010197790; 0010199877; 0010199878; 0010199880; 0010204029; 0010216175; 0010216176; 0010222009; 0010222011; 0010225047; 0010271291; 0010271292; 0010271293; 0010273071; 0010279050; 0010279938; 0010288574; 0010289862; 0010291595; 0010291597; 0010294650; 0010294651; 0010294652; 0010296300; 0010296301; Model Number/Description:D-EVPROP34US/DELIV SYS D-EVPROP34US SA GTIN:00763000207861 Lot Numbers: 0010165309; 0010165311; 0010291593; 0010294653; 0010294654; 0010303395;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2041-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2042-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2043-2021 | Class II | Ongoing | Voluntary: Firm initiated |