Z-0142-2022Class II

Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A

Canon Medical Systems USA, Inc. / initiated 2020-11-11 / Terminated / root cause: software

Canon Medical System, USA, INC. began a recall of Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A on . FDA classified it as Class II and gives the reason as "A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0142-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the diagnostic imaging system failing resulting in rescanning and reinjection of contrast medium.
Root cause tag
software
FDA root cause
Software design
Action
On November 11, 2020, Canon Medical Systems visited affected consignees and updated affected software. After completion of the software update, the consignee was provided a "Recall field Modification Instruction" document concerning the correction.
Quantity in commerce
3 units
Distribution
US Nationwide distribution in the states of NV, AR MN.
Product codes and lots
Affected serial numbers: 3DA1982001, 3DA2072002, 3DA2072003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0142-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0142-2022Class IITerminatedVoluntary: Firm initiated