Z-2207-2023Class II

Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3...

Medtronic Neurosurgery / initiated 2023-06-13 / Open, Classified / root cause: sterility

Medtronic Neurosurgery began a recall of Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106... on . FDA classified it as Class II and gives the reason as "Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may present clinically with signs and symptoms of an acute inflammatory...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-2207-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Neurosurgery
Product
Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DUREPAIR 5X8IN NCE, 62158.
Product code
GXQ Dura Substitute
Reason for recall
Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may present clinically with signs and symptoms of an acute inflammatory process, comparable to infection.
Root cause tag
sterility
FDA root cause
Process change control
Action
On 6/13/23 recall notices were mailed to customers and distributors and they were asked to do the following: 1) Return affected product to the recalling firm. 2) Complete and return the Customer Confirmation Form to [email protected] 3) This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Customers with questions can contact Customer Service at 800-874-5797 Option 1.
Quantity in commerce
65029
Distribution
Worldwide - US Nationwide distribution including in the states of PR, TX, WA, HI, MI, GA, AL, IL, IN, IA, MO, KY, NY, CA, AZ, PA, UT, ME, MD, MN, FL, AK, NE, VA, OR, WI, NC, LA, SD, TN, MS, DE, NJ, ID, SC, OH, MA, DC, NM, WV, NH, CT, NV, OK, AR, KS, MT, RI, CO, ND, WY, VT and the countries of Barbados, Cayman Islands, Honduras, Belgium, Switzerland, Italy, United Kingdom, Canary Islands, Hungary, Poland, Germany, Spain, Czech Republic, Thailand, South Africa, Greece, Finland, Portugal, Colombia, Norway, Iraq, Jordan, Estonia, Austria, Slovakia, Libya, Qatar, Bosnia And Herzegovina, Algeria, Dominican Republic, Canada, Australia, New Zealand, Sweden, El Salvador, Ireland, United Arab Emirates, Saudi Arabia, Kazakhstan, Turkey, Mauritius, Ghana, Russian Federation, Lebanon, Romania, Iran, Islamic Republic Of, Egypt, Namibia, Senegal, Ecuador, Panama, Peru, Guyana, Mexico, Chile, Costa Rica, Viet Nam, Kuwait, Ukraine, Israel, Turkmenistan.
Product codes and lots
REF/Expiration Date On or Prior/UDI-DI/Lots: 61100/March 31, 2025/00643169063808/2006039, 2007020, 2009028, 2010001, 2010016, 2102001, 2102020, 2103001, 2105028, 2106001, 2108011, 2108017, 2109022, 2110009, 2110030, 2111023, 2111045, 2112026, 2203020, 2204016, 2204035, 2207014, 2207046; 61105/February 28, 2025/00643169063815/2009060, 2011017, 2102011, 2102027, 2102047, 2103002, 2104042, 2105029, 2108001, 2108012, 2108016, 2109023, 2110010, 2111021, 2111024, 2111046, 2112027, 2202034, 2203034, 2204027, 2206008, 2207015, 2207033; 61106/December 31, 2024/00643169063822/2010017, 2103003, 2105030, 2105038, 2108013, 2108018, 2109004, 2110011, 2110031, 2111025, 2202035, 220302; 61110/March 31, 2025/00643169063839/2006039, 2010018, 2101014, 2102002, 2102028, 2103004, 2105031, 2108019, 2109024, 2110012, 2111026, 2201005, 2202036, 2204025, 2204040; 61111/April 30, 2025/00643169063846/2008039, 2010019, 2102021, 2103005, 2105032, 2106004, 2108014, 2110013, 2111027, 2202037, 2203022, 2203028, 2207016; 62100/January 31, 2025/00643169063853/2007021, 2008040, 2009026, 2009061, 2010014, 2011018, 2012001, 2101015, 2101020, 2102012, 2102022, 2102048, 2103019, 2104043, 2105033, 2106005, 2108020, 2109001, 2109020, 2109028, 2110014, 2110032, 2110033, 2111028, 2111035, 2111037, 2112001, 2112002, 2202038, 2203023, 2204001, 2204028, 2206013, 2206027, 2207017, 2207048, 2208009; 62105/January 31, 2025/00643169063860/2008005, 2008041, 2009029, 2010015, 2010020, 2012002, 2101016, 2102003, 2102013, 2102029, 2103006, 2103036, 2104044, 2105036, 2106035, 2108003, 2108015, 2109002, 2109003, 2110003, 2110015, 2110016, 2111022, 2111033, 2111047, 2201006, 2202014, 2202025, 2203015, 2203025, 2203031, 2204023, 2204036, 2204041, 2206014, 2206028, 2207018, 2207044; 62106/December 31, 2024/00643169063877/2008042, 2010021, 2101017, 2102023, 2103007, 2104045, 2105039, 2106007, 2108004, 2108021, 2109005, 2110017, 2110034, 2111029, 2202039, 2207019, 2207051; 62110/March 31, 2025/00643169063884/2006041, 2007022, 2008006, 2008043, 2010022, 2101009, 2101018, 2102026, 2102049, 2103020, 2104046, 2105001, 2105040, 2108005, 2108022, 2110004, 2110018, 2111036, 2201007, 2202015, 2202031, 2204017, 2204042, 2205031, 2205032, 2206007, 2207020, 2207049; 62111/May 31, 2025/00643169063891/2010023, 2012003, 2101019, 2102024, 2103008, 2105034, 2106009, 2108023, 2109006, 2110019, 2110035, 2111030, 2111034, 2202040, 2203024, 2204002, 2207021, 2207037, 2208027; 62158/April 30, 2025/00643169063907/2011019, 2103009, 2105035, 2110020, 2111031, 2202041, 2203033, 2204029, 2207006

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2207-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2207-2023Class IIOngoingVoluntary: Firm initiated