Citrine QTRAP Mass Spectrometer, Part Number 5063685
AB Sciex, LLC began a recall of Citrine QTRAP Mass Spectrometer, Part Number 5063685 on . FDA classified it as Class II and gives the reason as "In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0039-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- AB Sciex, LLC
- Product
- Citrine QTRAP Mass Spectrometer, Part Number 5063685
- Product code
- DOP Mass Spectrometer, Clinical Use
- Reason for recall
- In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- An URGENT: RUO and MEDICAL DEVICE FIELD ACTION notification letter dated 7/20/23 was sent to customers. Actions to be taken by the Customer/User: SCIEX will provide a replacement nut for eligible sources and systems. Instructions will be provided to guide you through the replacement of the nut. The replacement parts with installation instructions are estimated to begin shipping in October 2023. SCIEX is currently notifying affected customers of the upcoming countermeasure. Upon availability, SCIEX will send you the replacement nut with instructions on how to install it on the ion source probe. We sincerely apologize for any inconvenience that this may cause you. SCIEX aims to provide you with products of the highest quality. If you have any questions regarding this notice, please contact https://sciex.com/aboutus/ contact-us or https://sciex.com/support/request-support.
- Quantity in commerce
- 10
- Distribution
- US Nationwide.
- Product codes and lots
- UDI-DI: 00842365100540; Serial Numbers: DD250051901, DD250172009, DD250182009, DD230001804, DD250111910, DD250132004, DD250091909, DD250101910, DD250192010, CN10011707PT
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0037-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0038-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0039-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0040-2024 | Class II | Ongoing | Voluntary: Firm initiated |