MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504...

Canon Medical Systems USA, Inc. / initiated 2024-05-10 / Open, Classified / root cause: design

Canon Medical System, USA, INC. began a recall of MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010)... on . FDA classified it as Class II and gives the reason as "The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2041-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Canon Medical Systems USA, Inc.
Product
MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
Root cause tag
design
FDA root cause
Device Design
Action
On 5/10/24, correction notices were sent to customers who were asked to do the following: 1) The replacement of the power cable securing parts that may cause smoke and fire, with parts made of materials with a higher degree of flame retardance, is now being released. 2) A firm representative will contact you to schedule a time to replace the power cable securing parts. 3) If any abnormalities such as unusual smells or smoke are observed before the work, immediately stop using the system and contact your service representative. 4) The parts required to address the defect are expected to be available no earlier than June 2024. 5) Complete and return the reply form via DocuSign or via email to [email protected] For questions, please contact our InTouch team at [email protected]
Quantity in commerce
153
Distribution
US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR
Product codes and lots
Model/UDI-DI/Serial Numbers: MRT-1550 (MEXL-1550)/04987670103360/POA21X2002, S5A1982001, S5B22Z2041, S5B2172018, S2C2252024, S5A20Z2009, S2C21Y2021, S2B2052010, S5B2272031, POA2262005, S5B2292036, S5B2272029, S5B2282034, S5B22Y2040, S5B2292035, S6B2252008, S5A20Z2010, POA2252004, POA2182001, S1C2152071, S5B2352047, S2C2252023, S5B2122013, S6B2262010, S2C2192020, S6B2292013, S5B2132014, S6B2322018, S5B2292038, S6B2312016, S2B2052011, S2C2162018, S4A1992013, S3A19Z2013, S3A1982009, S5B2182021, S6B2322017, S5A2092008, S6B2362019, S5B22Z2042, S6B2222005, S2B2042009, S2B20Z2015, S5B2162017, S6B2262011, S5B2272030, S6B22Z2014, S5B2282033, S8B2272009, S5A2062007, S2B2072012, S2B2072013, S6B2222006, S6B2272012, S2B2072014, S6A20Z2001, POA2272006, S5A19Z2003, S1C2142070, POA2212003, S2C2142017, S5B2362048, S2A1962005, S6B2182004, S5B2292037, S6B2252007, S5B2282032, S5B2142015, S5B2382050, S2A18Z2003, S6B2112002, S6B2382021, S1B2042051, S8B2212007, S1B2062057, S1A1922022, S2A18Z2001, S1B2092067, S1C2242080, S5B2172019, S6B2252009, S6B2372020, S2A18Z2002, S2A1942004, S2C2212022, S2C2272025, S6B2392022, S2C2162019, S2C2132016, S4B2062020, S1C2262081, PDA21Y2001, S6B22Z2015, S5B22Z2043, S5B2242027, MRT-2020 (MEXL-1520)/04987670101632/S1F18X2580, S1B1492068, S1D17Y2461, S1F2052753, S1C1722390, S1C1682336, S1E1842517, S1E1842527, S1C1672334, MRT-1504 (MEXL-1504)/04987670102400/U5B15Z2001, MRT-3010 (MEXL-3010)/04987670101649/A5D14X2032, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5F1562034, A5B11Y2005, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5D1362021, A5C1252012, A5C1292015, A5D13X2024, A5C1422029, A5B11Z2007, A5C1212008, MRT-3020 (MEXL-3020)/04987670102615/D5A2082005, D5A2092007, D5B2152013, S5B17X2034, D7B2182017, 5DB2152002, S5B18X2046, S5A17X2033, D5A2062003, S5A1742021, D5B2212019, D5B2192016, S5A1662001, S5C18Y2048, S5A1732019, S5A1732018, S5A17Y2035, S7B1892004, S5B1842043, D7B2152015, S5C1972054, 5DB2172003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2041-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2041-2024Class IIOngoingVoluntary: Firm initiated