McKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0
CHANGE HEALTHCARE CANADA COMPANY began a recall of McKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0 on . FDA classified it as Class II and gives the reason as "A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0128-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Change Healthcare Canada Company
- Product
- McKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
- FDA root cause
- Use error
- Action
- Change Healthcare notified consignees on about 08/21/2024 via emailed letter. Consignees were instructed to pay careful attention to the hemoglobin unit of measurement displayed on the top of the "results" screen of the Hemo application, communicate the issue to all relevant team members to ensure vigilance in data entry, ensure that the hemoglobin values are entered in the correct unit of measurement (mmol/l or g/dl) as configured in the Management Console and displayed at the top of the results screen, contact Change Healthcare support to schedule the deployment of the software update that addresses this issue. Additionally, consignees were requested to complete and return the response form. With regard to potentially affected data, Change Healthcare recommended consignees verify f the UOM setting for hemoglobin is configured to what users would expect, if the UOM settings are different than expected, contact Change Healthcare support for assistance with identifying when the UOM setting was changed, and which procedures were affected, and examine affected procedures and implement necessary corrective actions to rectify the data and address patient care concerns.
- Quantity in commerce
- 35 systems
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom.
- Product codes and lots
- UDI Numbers: (01)80010939050010(10)130201, (01)80010939050027(10)140000. Versions Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0126-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0127-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0128-2025 | Class II | Ongoing | Voluntary: Firm initiated |