Z-0284-2025Class II

Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that...

Canon Medical Systems USA, Inc. / initiated 2024-09-17 / Open, Classified / root cause: software

Canon Medical System, USA, INC. began a recall of Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and... on . FDA classified it as Class II and gives the reason as "When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0284-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Canon Medical Systems USA, Inc.
Product
Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.
Root cause tag
software
FDA root cause
Software design
Action
On 9/17/24, correction notices were emailed to customers asking them to do the following: 1) The modified software will be installed to all applicable systems to correct the occurrence of this problem, which should be available 09/2024. 2) To avoid the problem, until the corrective actions have been taken, it is strongly recommended that queue of more than one reconstruction be avoided. 3) If PET reconstruction stops unexpectedly, perform the following steps: <1> Execute "Reset PET System" and reboot the PET system. <2> Delete the remaining jobs in the reconstruction queue. <3> After the PET system has rebooted, start up raw data processing and select the appropriate data. <4> Perform one of the following from the pull-down menu for "Helical for AC". . Specify the helical scan line number to perform CT reconstruction. . Specify "Select Image" and select the CT data used for attenuation correction. <5> Perform PET reconstruction. If PET reconstruction cannot be performed because the CT data for attenuation correction is not saved or for any other reason, follow the procedure below. Specifying helical scan data used for attenuation correction. For helical + PET acquisition, the helical scan data used for attenuation correction can be specified. When "Original" is specified, the CT data used for reconstruction synchronized with scanning is used. When the helical scan line number is specified, reconstruction for helical scan is performed first. When "Select Image" is specified, the data used for attenuation correction can be selected from the displayed list, see Image Reconstruction in the operation manual. 4) Share this information with all users, radiologists, clinical engineering, or biomedical groups. 5) Complete and return the response form via email to [email protected] 6) If you have questions please email them to the address above.
Quantity in commerce
27
Distribution
US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA
Product codes and lots
UDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0284-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0284-2025Class IIOngoingVoluntary: Firm initiated