Z-0327-2026Class II

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7

Canon Medical Systems USA, Inc. / initiated 2025-10-16 / Open, Classified / root cause: design

Canon Medical System, USA, INC. began a recall of VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7 on . FDA classified it as Class II and gives the reason as "Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0327-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Canon Medical Systems USA, Inc.
Product
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Root cause tag
design
FDA root cause
Device Design
Action
Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form
Quantity in commerce
2 units
Distribution
US distribution to Ohio and Nevada
Product codes and lots
CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0327-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0327-2026Class IIOngoingVoluntary: Firm initiated