Z-0327-2026Class II
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7
Canon Medical Systems USA, Inc. / initiated 2025-10-16 / Open, Classified / root cause: design
Canon Medical System, USA, INC. began a recall of VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7 on . FDA classified it as Class II and gives the reason as "Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0327-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form
- Quantity in commerce
- 2 units
- Distribution
- US distribution to Ohio and Nevada
- Product codes and lots
- CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0327-2026 | Class II | Ongoing | Voluntary: Firm initiated |