Alphenix INFX-8000C, interventional fluoroscopic x-ray system
Canon Medical System, USA, INC. began a recall of Alphenix INFX-8000C, interventional fluoroscopic x-ray system on . FDA classified it as Class II and gives the reason as "It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1176-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Alphenix INFX-8000C, interventional fluoroscopic x-ray system
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Canon Medical Systems USA issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2025 via letter. The notice explained the issue, risk to health, and requested the following: "Request to customers: Should the suspended device make an abnormal noise, or a sensor error is displayed before the corrective action is implemented on your system, please stop using the device and call your local Canon Service Team. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at [email protected]."
- Quantity in commerce
- 172 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Dominican Republic.
- Product codes and lots
- serial numbers: W1C0632284, W4A1282279, W4A12Z2324, A2A2222381, A2A2392756, A3517001, A2A2352683, A2A2072087, A2A2072077, C1C2422647, W2A09Y2061, W2A10X2168, W4A11Y2144, A2A2232406, A2A2032038, W1D07Z2458, W2A09X2053, B2B1812645, A1C19Z2182, W4B1332362, B2A1622324, W4C1352385, W4A1292284, A1C19X2152, A1C1942058, A1C1972104, B2A1592233, B2A1792598, W2A1012081, W2A1052121, W4C1342374, B2B1872766, A4562105, W2A0972027, A1C2052199, A2A21X2339, W2A1012082, W4C1342369, W1D07Z2449, A1C1942065, A2A2132228, A2A2172285, A2A2117001, W1D06Z2357, A2A2282477, A2A2232405, W1B08X2034, A2A2182313, W4A1232210, A2A2262452, B2A16X2449, B2B1862743, B2B1872751, W1D0697014, W1C05X2229, B2A1542162, A2A21Z2364, A2A2392751, B2A1722527, A2A22Z2566, B4582115, W4C1372400, A2A2362690, A2A2252436, A2A2292510, B2A1582209, B2A1472014, W1D0732376, A2A2252439, W1D0732374, A1C1972105, A1C1972112, B4602126, W1B08Z2063, A2A2152265, A2A2132230, A2A2282491, A2A2262445, A2A22Y2550, A2A22X2528, W4A1212180, W4A1212176, W4A1212177, A1C1992141, W4A1182110, W4A11Z2163, W1D0812466, B2B18Y2816, B2B1832693, A1C1952078, A2A2062070, W2A09Z2077, A1C19Y2174, W1D0722372, A2A22X2518, A3542008, B2A1472020, W4A1242224, W4A1312348, A2A2172280, W4D1422466, W4A1142069, A2A2172287, A2A2342657, A2A2432872, B2A1552176, B2B1812657, B2B1872765, W4A1342373, B2A1682407, W4A11Z2170, W4C1342364, W4C1342365, W4A1172092, A2A2172281, W2A1072143, W4A1292281, A2A2242426, A2A2392761, A3612061, A1A2162240, A2A23Z2814, B2B1872760, A2A2262456, A1C2042197, B2A14Z2085, W4C1352377, W4A1252245, A1C19Z2178, A2A2292499, W4A1242221, W1D0872524, A1C19Z2188, A1C19X2158, A2A2052053, B2B1862731, B2A1752550, B2B1832678, B2B1852708, B2B1852710, B2B1852707, W1C0582207, W1C05Y2242, W1D0722370, W4A1272267, W2A1062127, B2A1742539, B2B1922842, A2A2062066, B2A16Y2469, W4A1122048, B2A1662369, W4A1122039, B5512160, A2A2122205, B2B1822664, B2B1822670, A2A2292498, W4D13Y2442, A2A2122197, A2A2192330, A2A2132216, A2A22X2522, A2A2432885, A2A2422855, W2A1012091, W1C05X2224, B2A1622310, A1C1972111, B2B1892787, A1C1952080, A1B18Y7002.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1176-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1177-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1178-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1179-2026 | Class II | Ongoing | Voluntary: Firm initiated |