Z-2624-2026Class I

Medtronic Harmony Delivery Catheter System

Medtronic Heart Valves Division / initiated 2026-05-22 / Open, Classified / root cause: manufacturing process

Medtronic Heart Valves Division began a recall of Medtronic Harmony Delivery Catheter System on . FDA classified it as Class I and gives the reason as "Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2624-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Heart Valves Division
Product
Medtronic Harmony Delivery Catheter System
Product code
NPV Pulmonary Valve Prosthesis Percutaneously Delivered
Reason for recall
Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Medtronic issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 5/28/2026 via mail or email. The notice explained the issue, potential risk, and requested the following: " - Review your inventory for affected lot numbers listed. - If unused inventory from the listed lot numbers is located, immediately quarantine and return the affected units to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of the affected units. Your Medtronic sales representative can assist in the return of the affected units as necessary. - Please share this notification with implanters and teams within your organization. If product listed above has been forwarded from your facility to another facility, please notify the other facility of this Medtronic Urgent Medical Device Recall. - Complete the enclosed Customer Confirmation Form and email to [email protected]. This form must be returned even if you do not have the affected units in your possession. - Please maintain a copy of this communication in your records."
Quantity in commerce
1881 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Canada, Chile, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Mexico, Netherlands, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Product codes and lots
GTIN 00763000341367, Lot Numbers: 0012508285, 0012517045, 0012526102, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663878, 0012663879, 0012690218, 0012699635, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824770, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965210, 0012965211, 0012976400, 0012986693, 0013063766, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624, 0013236643, 0013272796, 0013272798. GTIN 00763000544027, Lot Numbers: 0012508285, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663879, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965211, 0012976400, 0012986693, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624. GTIN 00763000520151, Lot Numbers: 0012575897, 0012642965, 0012728368, 0013272799. GTIN 00763000582951, Lot Numbers: 0012590535, 0012606411, 0012606413, 0012642964, 0012659418, 0012678282, 0012690219, 0012699637, 0012804151, 0012804580, 0012874527, 0012888682, 0012896885, 0012965221, 0012986699, 0013063764, 0013063765, 0013105037, 0013227504. GTIN 00763000918712, Lot Numbers: 0012635392, 0012676200, 0012690207, 0012874528, 0012874529, 0013015539, 0013105038.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2624-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2624-2026Class IOngoingVoluntary: Firm initiated