Z-0061-05
Excelart (tm) VANTAGE P2 & P3, Magnetic Resonance Imaging System
Canon Medical Systems USA, Inc. / initiated 2004-09-16 / Terminated / root cause: software
Toshiba American Med Sys Inc began a recall of Excelart (tm) VANTAGE P2 & P3, Magnetic Resonance Imaging System on . The reason FDA records is "Scanned images are acquired at a position that is shifted relative to the intended position in the slice direction due to a software anomaly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0061-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Excelart (tm) VANTAGE P2 & P3, Magnetic Resonance Imaging System,
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Scanned images are acquired at a position that is shifted relative to the intended position in the slice direction due to a software anomaly.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Firm issued a letter to users on 9/29/2004 which includes a method to avoid the software problem until it can be corrected by an update.
- Quantity in commerce
- 22
- Distribution
- PA, MD, FL, OH, NY, CA, AR, KY, OK, AZ, TX, NV, IN.
- Product codes and lots
- Serial numbers: A4512006, A452017, A4532011, A3612004, A4532014, A4572020, A4572021, A3622005, A4562003, A3592003, A4562018, A4522008, A4532010, A4542015, A4532013, A3622001, A3582001, A4522009, A4532012, A3592002, A4522007, A4552016.