Z-0061-05

Excelart (tm) VANTAGE P2 & P3, Magnetic Resonance Imaging System

Canon Medical Systems USA, Inc. / initiated 2004-09-16 / Terminated / root cause: software

Toshiba American Med Sys Inc began a recall of Excelart (tm) VANTAGE P2 & P3, Magnetic Resonance Imaging System on . The reason FDA records is "Scanned images are acquired at a position that is shifted relative to the intended position in the slice direction due to a software anomaly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0061-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Excelart (tm) VANTAGE P2 & P3, Magnetic Resonance Imaging System,
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Scanned images are acquired at a position that is shifted relative to the intended position in the slice direction due to a software anomaly.
Root cause tag
software
FDA root cause
Other
Action
Firm issued a letter to users on 9/29/2004 which includes a method to avoid the software problem until it can be corrected by an update.
Quantity in commerce
22
Distribution
PA, MD, FL, OH, NY, CA, AR, KY, OK, AZ, TX, NV, IN.
Product codes and lots
Serial numbers: A4512006, A452017, A4532011, A3612004, A4532014, A4572020, A4572021, A3622005, A4562003, A3592003, A4562018, A4522008, A4532010, A4542015, A4532013, A3622001, A3582001, A4522009, A4532012, A3592002, A4522007, A4552016.