Z-0270-05

Magnetic Resonance Imaging Systems under the following names: Visart Excelart Excelart P2 Excelart P3...

Canon Medical Systems USA, Inc. / initiated 2004-10-06 / Terminated / root cause: software

Toshiba American Med Sys Inc began a recall of Magnetic Resonance Imaging Systems under the following names: Visart Excelart Excelart P2 Excelart P3 Excelart SPIN Excelart VANTAGE on . The reason FDA records is "Software defect causes images to be rotated 180 degrees". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0270-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Magnetic Resonance Imaging Systems under the following names: Visart Excelart Excelart P2 Excelart P3 Excelart SPIN Excelart VANTAGE
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Software defect causes images to be rotated 180 degrees.
Root cause tag
software
FDA root cause
Other
Action
Firm sent a letter to consignees October 6, 2004 describing the observed conditions causing the flaw to manifest itself. A workaround solution is also described in the letter. A software correction is under development and will be installed by a TAMS representative in the future.
Quantity in commerce
67
Distribution
Nationwide
Product codes and lots
All serial numbers.