Z-0273-05
Magnetic Resonance Imaging Systems, under the following names: Visart Excelart Excelart P2 Excelart P3...
Canon Medical Systems USA, Inc. / initiated 2004-08-24 / Terminated / root cause: software
Toshiba American Med Sys Inc began a recall of Magnetic Resonance Imaging Systems, under the following names: Visart Excelart Excelart P2 Excelart P3 Excelart SPIN on . The reason FDA records is "Software defect causes patient identification number to be truncated". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0273-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Magnetic Resonance Imaging Systems, under the following names: Visart Excelart Excelart P2 Excelart P3 Excelart SPIN
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Software defect causes patient identification number to be truncated.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Firm issued a letter notifying users of systems that when transferring image data to a DICOM server for any patient born on January 1, to verify that the Patient ID is correct on the DICOM server after the transfer. Firm will modify the software by January 2005.
- Quantity in commerce
- 39 units
- Distribution
- Nationwide.
- Product codes and lots
- No specific codes were listed.