Z-0982-05

Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002

Canon Medical Systems USA, Inc. / initiated 2005-05-23 / Terminated

Toshiba American Med Sys Inc began a recall of Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002 on . The reason FDA records is "If the measurement mode is started from the Doppler mode and then the sweep speed setting is changed by the assigned user function, the measurement of the time direction results don't correspond to the actual values". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0982-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
If the measurement mode is started from the Doppler mode and then the sweep speed setting is changed by the assigned user function, the measurement of the time direction results don't correspond to the actual values.
FDA root cause
Other
Action
Firm issued a recall letter which provided a procedural workaround, a software patch will be installed during routine service visits.
Quantity in commerce
106
Distribution
Nationwide. This is an imported product.
Product codes and lots
Software versions 6.0er001 and 002