Z-0982-05
Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002
Canon Medical Systems USA, Inc. / initiated 2005-05-23 / Terminated
Toshiba American Med Sys Inc began a recall of Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002 on . The reason FDA records is "If the measurement mode is started from the Doppler mode and then the sweep speed setting is changed by the assigned user function, the measurement of the time direction results don't correspond to the actual values". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0982-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- If the measurement mode is started from the Doppler mode and then the sweep speed setting is changed by the assigned user function, the measurement of the time direction results don't correspond to the actual values.
- FDA root cause
- Other
- Action
- Firm issued a recall letter which provided a procedural workaround, a software patch will be installed during routine service visits.
- Quantity in commerce
- 106
- Distribution
- Nationwide. This is an imported product.
- Product codes and lots
- Software versions 6.0er001 and 002