Z-0986-05
Magnetic Resonance Imaging Systems; Visart, Excelart / Excelart P2 / Excelart P3 / Excelart SPIN & Excelart...
Canon Medical Systems USA, Inc. / initiated 2005-05-25 / Terminated
Toshiba American Med Sys Inc began a recall of Magnetic Resonance Imaging Systems; Visart, Excelart / Excelart P2 / Excelart P3 / Excelart SPIN & Excelart VANTAGE on . The reason FDA records is "Patient information A is displayed with images for patient B". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0986-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Magnetic Resonance Imaging Systems; Visart, Excelart / Excelart P2 / Excelart P3 / Excelart SPIN & Excelart VANTAGE.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Patient information A is displayed with images for patient B.
- FDA root cause
- Other
- Action
- Firm will have technicians correct each unit by visit. Instructions to them have issued.
- Quantity in commerce
- 102
- Distribution
- Nationwide.
- Product codes and lots
- All.