Z-1170-06
62 RNR Series 2.5 microliter syringe without needle, Part#: 7632-01
Hamilton Co / initiated 2006-03-28 / Terminated
Hamilton Co began a recall of 62 RNR Series 2.5 microliter syringe without needle, Part#: 7632-01 on . The reason FDA records is "There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1170-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hamilton Co
- Product
- 62 RNR Series 2.5 microliter syringe without needle, Part#: 7632-01
- Product code
- JQW Station, Pipetting And Diluting, For Clinical Use
- Reason for recall
- There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe
- FDA root cause
- Other
- Action
- On 3/28/06, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.
- Quantity in commerce
- 36 units
- Distribution
- Worldwide-The product was distributed to 28 U.S. consignees and 3 foreign consignees. (Canada, Switzerland, Japan)
- Product codes and lots
- Lot 342268 and 345852