Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR

Hamilton Co

Reno, NV, US

HAMILTON CO. appears in FDA device records in Reno, NV. FDA records list 16 510(k) clearances between and , no PMA records and one registered establishment. As of , FDA records show 13 recalls for this company, and at least one establishment registration is current.

FDA records held for this company

510(k) clearances
16
PMA records
0
Recall records
13
Registered establishments
1

510(k) clearance history

FDA records hold 16 510(k) clearances for Hamilton Co between 1980 and 1996. The busiest year on record is 1983, with 9. The most recent is 1996, with 1.

Clearances by year, as a table
510(k) clearance decisions for Hamilton Co, by decision year
YearClearances
19801
19821
19839
19871
19881
19891
19951
19961
510(k) clearances on record, most recent first
K numberDeviceProduct codeDecisionDecision date
K961896MICROLAB 4000, MICROLAB 4200JQWSubstantially Equivalent
K943493DISCRETE PHOTOMETRIC CHEMISTRY ANALYZERJJESubstantially Equivalent
K892369MULTI-PROBE HEAD PLUSJTCSubstantially Equivalent
K873893LUMICONJJQSubstantially Equivalent
K871945MICROLAB ATJTCSubstantially Equivalent
K832205PRECISION LIQUID DISPENSERJQWSubstantially Equivalent
K832204MICROLAB PJQWSubstantially Equivalent
K832202PB 600 REPEATING DISPENSERJQWSubstantially Equivalent
K832201CHROM-PREPKZRSubstantially Equivalent
K832200PRP-1 REVERSED PHASE COLUMNS FOR HPLCDPMSubstantially Equivalent
K832199HC40/HC75 CATION EXCHANGE RESINDNHSubstantially Equivalent
K832198HC40/HC75 PREPACKED COLUMSDJYSubstantially Equivalent
K832197PRP-1 REVERSED PHASE RESINDMZSubstantially Equivalent
K832203MULTI-TITER DISPENSERJTCSubstantially Equivalent
K822688DIGITAL DILUTER/DISPENSERSJQWSubstantially Equivalent
K801913MICROLAB M-MICROPROCESSOR DILUTER/DISPJQWSubstantially Equivalent

Registered establishments

FDA registered establishments for Hamilton Co, with the product codes listed at each
RegistrationFEIEstablishmentAddressRegistration expiresProduct codes listed
29157962915796HAMILTON CO.4970 ENERGY WAY, Reno, NV 89502 US2026

Registration numbers link to the FDA registration and listing record for that establishment.

FDA product codes

Recall records

Device recall records naming this company, most recent first
RecallProductReasonClassificationInitiatedStatus
Z-1156-2021HAMILTON HEATER SHAKER, 115 VAC/ 230 VAC, 750 VA, Power Consumption 41V/ 100W (max.), Relative Humidity 15% - 85% no condensation. REF 19903400There is a risk that the device can start a fire if flammable liquids contact and enter the device.Class IIOpen, Classified
Z-0103-2012Hamilton brand Disposable Syringe Pack for use with Microlab F.A.M.E; Part Number: 147692 (case containing 96 syringes); Manufactured by: Treff AG, Degersheim, Switzerland for Hamilton Bonaduz AG, Bonaduz, Switzerland (Manufacturer of record); Product is distributed by Hamilton Company, Reno, NV The F.A.M.E Syringes are a disposable product that is used on our Microlab F.A.M.E. series instruments for diluting/dispensing. Microlab F.A.M.E was released under 510(k) K943493/S1 under product codes JJE (862.2160 Discrete Photometric Chemistry Analyzer for Clinical Use), JJQ (862.2300 Colorimeter, photometer, or spectrophotometer for Clinical Use), and JTC (866.2500 Microtiter Diluting/Dispensing Device), all of which are currently Class I exempt devices. The syringes are also used on Ortho Summit Processor instruments distributed by Ortho-Clinical Diagnostics.Certain lots of syringes have caused an increase in error message "DIS541", which indicates "improper dispensation".Terminated
Z-1999-2009300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. Product is a disposable product used with Microlab STAR series Instruments, which are microtiter diluting and dispensing devices.Product Defect-- some tips from the mold cavity 28 are too short and do not always aspirate or dispense liquid correctly.Terminated
Z-1203-2009300 microliter CO-RE tips for use with Microlab STAR series instruments, Part Number 235902, 5,760 tips per case, manufactured by Hamilton Company, Reno, NV. The products are a disposable product that are used with the Microlab STAR series instruments, which are microtiter diluting and dispensing devices.Some tips may be occluded and are unable to aspirate or dispense liquid.Terminated
Z-2233-2008Hamilton Soft Grip Pipettes, 1 ml adjustable volume, model number 55019-40 Product is indicated for manual pipetting.Manufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacityTerminated
Z-2232-2008Hamilton Soft Grip Pipettes, 1 ml fixed volume, model number 55019-19 Product is indicated for manual pipetting.Manufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacityTerminated
Z-0718-2008Hamilton Sample Carriers for the ML STAR line (SMP-CAR-24). Product Code: 866.2500; Part Number: 173400. Hamilton Company, Reno, NV 89520 (Accessory on the Microlab STAR series instruments.)Incorrectly labeled-- Product has the wrong barcode labeling. The barcode error will not influence the results of testing. The instrument will discover the error and post an error notice.Terminated
Z-1169-2007Microlab F.A.M.E. reagent containers, Part number 148482, box containing 48 containers each. For in vitro diagnostic use.Some product containers may be leaking.Terminated
Z-0723-2007Hamilton brand Disposable Precision Tips, Part Number 235300 (box containing 504 tips) and 235400 (case containing 20 boxes), Manufactured by Hamilton Company, Reno, NevadaSome of the packaged tips have increased force during aspiration, which can lead to pipetting errors such as LLD error messages, and also has risk of causing a missed pipetting.Terminated
Z-1170-0662 RNR Series 2.5 microliter syringe without needle, Part#: 7632-01There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringeTerminated
Z-1169-06Syringe, 62RNR 2.5 Microliter Syringe, with needle point style 3 (blunt needle point for use with HPLC injection valves and for sample pipetting) Part #: 87942.There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringeTerminated
Z-0514-06Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008, 135016, 135028, 135029, 936400, 936400BLU. Microlab AT+ and Microlab AT+2, part numbers 142000, 142001, 142001R, 142002, 142003, 142005, 142006, 142006R, 142007, 142008, 142009, 142015, 142017, 142035, 142036Heat sinking on power boards may not operate as intended and the board may cease to operate.Terminated
Z-0087-06Y-valve and T-valve, PSD/4 Syringe Drive Module parts.The product valve has a high probability of deforming and may result in leakage during use.Terminated

Verification record

There is no verification record for this company. Most companies on this site have none: a record exists only where a company asked to be verified and MDM ran the checks, so the absence of one says nothing about the company.

Where there is a record, this is where it appears: what was checked, when it was checked, and the current status. An MDM Verified badge on another website links to a verify address on this domain that resolves to this section. A badge that does not resolve here is not ours.

What MDM checks, and what it does not

MDM coverage

MDM coverage of this company lists here, newest first. Submit news.

Is this your company?

This page is built from public FDA records. Company pages are free and factual, and stay free whether or not anyone claims them. Membership never changes editorial coverage or vetting outcomes.

Claiming lets Hamilton Co edit what it says about itself and list the categories it works in. It grants no badge: claiming is not verification.

Claim this page

Data as of , FDA export datePage updated Source: openFDA, public domain