Z-1497-06
Skytron General Purpose Surgical Table, battery model; Model 6600B, Dist. by Skytron, Grand Rapids, MI
Skytron, LLC / initiated 2005-08-04 / Terminated
Skytron, Div. The KMW Group, Inc began a recall of Skytron General Purpose Surgical Table, battery model; Model 6600B, Dist. by Skytron, Grand Rapids, MI on . The reason FDA records is "The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1497-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Skytron, LLC
- Product
- Skytron General Purpose Surgical Table, battery model; Model 6600B, Dist. by Skytron, Grand Rapids, MI.
- Product code
- FQO Table, Operating-Room, Ac-Powered
- Reason for recall
- The release levers can inadvertently release under load while the leg section is positioned at a ninety degree angle causing the leg section to drop.
- FDA root cause
- Other
- Action
- A service bulletin was issued to distributors on 8/12/05 instructing them to identify customers and to retrofit these tables.
- Quantity in commerce
- 160 total
- Distribution
- Nationwide.
- Product codes and lots
- All units. Serial numbers 6600B-2D-002 through 6600B-2L-060.