Z-0534-2007
Skytron Infinity Series surgical light with triple lightheads, 8 bulbs in 1 - 30' diameter lighthead and 5...
Skytron, LLC / initiated 2007-01-08 / Terminated
Skytron, Div. The KMW Group, Inc began a recall of Skytron Infinity Series surgical light with triple lightheads, 8 bulbs in 1 - 30' diameter lighthead and 5 bulbs in each of 2 - 22' diameter lightheads... on . The reason FDA records is "If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0534-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Skytron, LLC
- Product
- Skytron Infinity Series surgical light with triple lightheads, 8 bulbs in 1 - 30' diameter lighthead and 5 bulbs in each of 2 - 22' diameter lightheads, Manufactured by Dai-ichi Shomei Co., Ltd., Made in Japan; Model IN302222EL, IF302222-B, IN30AR2222EL and IN30TV2222EL. (note only the 22'' lightheads are affected).
- Product code
- FTD Lamp, Surgical
- Reason for recall
- If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.
- FDA root cause
- Other
- Action
- The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.
- Quantity in commerce
- 5,000 for all models.
- Distribution
- Nationwide and Canada.
- Product codes and lots
- Each lighthead bears a unique serial number. For units with model numbers containing IF and 22, all serial numbers with the following numbers in the third and fourth position of the numerical series: 95, 96, 97, 98, 99 or 00 (examples: xx97xxxx or xx00xxxB). For units with model numbers containing IN and 22 , all serial numbers with the following numbers in the third and fourth positions of the numerical series: 92, 93 or 94 (example: xx93xxxx).