Z-0771-2007

Nemio XG, ultrasound system, SSA-590A with version 1.5 software, Toshiba American Medical Systems, Inc...

Canon Medical Systems USA, Inc. / initiated 2006-12-19 / Terminated

Toshiba American Med Sys Inc began a recall of Nemio XG, ultrasound system, SSA-590A with version 1.5 software, Toshiba American Medical Systems, Inc., Tusin, CA on . The reason FDA records is "Transducer Surface Temperature : It was found during testing that when using the system in the 2DCDI + PW mode, with the Scale Graph function active, that the transducer surface temperature could exceed the 27 degrees...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0771-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Nemio XG, ultrasound system, SSA-590A with version 1.5 software, Toshiba American Medical Systems, Inc., Tusin, CA.
Product code
ITX Transducer, Ultrasonic, Diagnostic
Reason for recall
Transducer Surface Temperature : It was found during testing that when using the system in the 2DCDI + PW mode, with the Scale Graph function active, that the transducer surface temperature could exceed the 27 degrees Celsius limit as specified in the IEC safety standard.
FDA root cause
Other
Action
Notification letters were sent to customers on December 19, 2206. Customers were instructed to avoid using Scale Graph function with live images, and instructed to call up the Scale Graph after images are frozen. Customers were informed that a Toshiba Customer Service Engineer will contact them to schedule a software upgrade when software becomes available. If customers have any questions they were instructed to contact their Toshiba Customer Service Engineer or call (800) 421-1968 and ask for the Regulatory Affairs Department. Customers were also instructed to distribute this information to all users of the product and also with the Bio-medical Department at their facility.
Quantity in commerce
37
Distribution
Nationwide Distribution including the states of CA, FL, GA, ID, MA, MN, NC, NY, PA, RI, TN, TX, WA, and WI.
Product codes and lots
Serial #: A1A0642100; A1A0642102; A1B0662183; A1B0662184; A1B0662256; A1B0662257; A1B0662258; A1B0662327; A1B0662328; A1B0662330; A1B0662331; A3A0652115; A3B0662284; A3B0662285; A3B0662286; A3B0662287; A3B0662288; A3B0662289; A3B0662290; A3B0662291; A3B0662292; A3B0662293; A3B0662371; A3B0662372; A3B0662373; A3B0662374; A3B0662375; A3B0662376; A3B0662377; A3B0662378; A3B0662379; A3B0662380; A3B0672415; A3B0672416; A3B0672417; A3B0672418; A3B0672419;