Hamilton Sample Carriers for the ML STAR line (SMP-CAR-24)
Hamilton Co began a recall of Hamilton Sample Carriers for the ML STAR line (SMP-CAR-24). Product Code: 866.2500; Part Number: 173400. Hamilton Company, Reno, NV 89520 (Accessory on the... on . The reason FDA records is "Incorrectly labeled-- Product has the wrong barcode labeling. The barcode error will not influence the results of testing. The instrument will discover the error and post an error notice". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0718-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hamilton Co
- Product
- Hamilton Sample Carriers for the ML STAR line (SMP-CAR-24). Product Code: 866.2500; Part Number: 173400. Hamilton Company, Reno, NV 89520 (Accessory on the Microlab STAR series instruments.)
- Product code
- JTC Device, Microtiter Diluting/Dispensing
- Reason for recall
- Incorrectly labeled-- Product has the wrong barcode labeling. The barcode error will not influence the results of testing. The instrument will discover the error and post an error notice.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- The firm sent an Urgent: Product Recall letter to consignees on June 25, 2007. The firm also issued an alert via their User news distribution. They instructed consignees to examine their stock and return any affected carriers. A fax back form was supplied with the notification. The firm will be providing replacements.
- Quantity in commerce
- 179 units distributed.
- Distribution
- Worldwide Distribution --- USA including states of OH, CA, NJ, NC, and countries of Japan, France, Switzerland, Spain, Italy, Germany, and the UK.
- Product codes and lots
- Serial Numbers: S0213281 to S0214480.