Z-1707-2008
Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system
Canon Medical Systems USA, Inc. / initiated 2007-10-01 / Terminated / root cause: design
Toshiba American Med Sys Inc began a recall of Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system on . The reason FDA records is "Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1707-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Toshiba sent a letter to customers in July 2007, and sent field service representatives to update the Data Aquisition System in September/October 2007.
- Quantity in commerce
- 1203 total
- Distribution
- Worldwide distribution, including USA and Japan.
- Product codes and lots
- All codes