Z-1707-2008

Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system

Canon Medical Systems USA, Inc. / initiated 2007-10-01 / Terminated / root cause: design

Toshiba American Med Sys Inc began a recall of Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system on . The reason FDA records is "Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1707-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Systems USA, Inc.
Product
Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used.
Root cause tag
design
FDA root cause
Device Design
Action
Toshiba sent a letter to customers in July 2007, and sent field service representatives to update the Data Aquisition System in September/October 2007.
Quantity in commerce
1203 total
Distribution
Worldwide distribution, including USA and Japan.
Product codes and lots
All codes