Z-1139-2009

Snap Shunt Ventricular Catheter, Standard, with BioGlide, 8 CM, Ref 27708E. Each catheter is provided in a...

Medtronic Neurosurgery / initiated 2009-02-12 / Terminated / root cause: design

Medtronic Neurosurgery began a recall of Snap Shunt Ventricular Catheter, Standard, with BioGlide, 8 CM, Ref 27708E. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed... on . The reason FDA records is "The ventricular catheters may become detached from the Snap Base assembly after implantation and this condition may increase the need for emergency revision surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1139-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neurosurgery
Product
Snap Shunt Ventricular Catheter, Standard, with BioGlide, 8 CM, Ref 27708E. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The sealed pouch is then placed within a chipboard carton along with product IFU and patient wallet cards. The product is ETO sterilized.
Product code
JXG Shunt, Central Nervous System And Components
Reason for recall
The ventricular catheters may become detached from the Snap Base assembly after implantation and this condition may increase the need for emergency revision surgery.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent - Medical Device Recall notification was sent on February 13,2009 to the attention of risk managers, advising them to cease use of the Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782 ( all versions and all lot numbers), the Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27802 ( all versions and all lot numbers), and the Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27708 ( all version and all lot numbers). Each customer is to cease the use of the listed product and any unused unit needs to be returned to Medtronic. Further, each account is required to account for product used/implanted and to advise surgeons of the recall information. Each account will recieved an individualized Customer Product Reconciliation Form provided by their Medtronic Sales Representative for determining the status of any product shipped to them.
Quantity in commerce
3048 of all units
Distribution
Worldwide Distribution -- USA, Ireland, and Australia.
Product codes and lots
All Lots