Z-2339-2010
Toshiba Aquilion ONE CT SystemTSX-301 A/2D
Canon Medical Systems USA, Inc. / initiated 2009-02-03 / Terminated / root cause: software
Toshiba American Medical Systems Inc began a recall of Toshiba Aquilion ONE CT SystemTSX-301 A/2D on . The reason FDA records is "ECG Gated Dynamic Volume Prospective scan may stop if preset eXam Plan is confirmed and CTA On is changed to CTA OFF, due to exposure control PWB". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2339-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Medical Systems USA, Inc.
- Product
- Toshiba Aquilion ONE CT SystemTSX-301 A/2D
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- ECG Gated Dynamic Volume Prospective scan may stop if preset eXam Plan is confirmed and CTA On is changed to CTA OFF, due to exposure control PWB.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Toshiba America Medical System Inc (TAMS) sent an Urgent: Medical Device Correction dated January 14, 2009 to their consignees. The letter described the issue and provided instructions to test the device to determine if it is an affected device. If the device is affected, contact the Toshiba service representative at 800 521-1968. Consignees were requested to complete and fax a response form to Toshiba.
- Quantity in commerce
- 16
- Distribution
- Nationwide Distribution: including the states of AL, CA, FL, GA, KY, MA, MD, MT, NV, NY, OH, PA, and Puerto Rico.
- Product codes and lots
- 2DA0872038, 2DA0862034, 2DA07X2008, 2DA0842023, 2DA07X2005, 2DA0882053, 2DA0882051, 2DA0862036, 2DA0882049, 2DA0882052, 2DA0842026, 2DA07Y2010, 2DA0882048, 2DA08X2061, 2DA0792004, and 2DA0792002